Research Letter
Abstract
Background: The growing global burden of visual impairment necessitates better population eye screening for early detection of eye diseases. However, accessibility to testing is often limited and centralized at in-hospital settings. Furthermore, many eye screening programs were disrupted by the COVID-19 pandemic, presenting an urgent need for out-of-hospital solutions.
Objective: This study investigates the performance of a novel remote perimetry application designed in a virtual reality metaverse environment to enable functional testing in community-based and primary care settings.
Methods: This was a prospective observational study investigating the performance of a novel remote perimetry solution in comparison with the gold standard Humphrey visual field (HVF) perimeter. Subjects received a comprehensive ophthalmologic assessment, HVF perimetry, and remote perimetry testing. The primary outcome measure was the agreement in the classification of overall perimetry result normality by the HVF (Swedish interactive threshold algorithm–fast) and testing with the novel algorithm. Secondary outcome measures included concordance of individual testing points and perimetry topographic maps.
Results: We recruited 10 subjects with an average age of 59.6 (range 28-81) years. Of these, 7 (70%) were male and 3 (30%) were female. The agreement in the classification of overall perimetry results was high (9/10, 90%). The pointwise concordance in the automated classification of individual test points was 83.3% (8.2%; range 75%-100%). In addition, there was good perimetry topographic concordance with the HVF in all subjects.
Conclusions: Remote perimetry in a metaverse environment had good concordance with gold standard perimetry using the HVF and could potentially avail functional eye screening in out-of-hospital settings.
doi:10.2196/45044
Keywords
Introduction
Population eye screening programs are critical for early detection of eye diseases to enable timely interventions and address preventable blindness [
, ]. The COVID-19 outbreak has resulted in massive disruptions to health care services [ , ], including the postponement of health care procedures as well as reduced patient uptake of elective services due to fears of contracting COVID-19 infection [ ]. Despite the many successes in the continuity of care enabled by technology during the COVID outbreak, its disruptive impact still remains, particularly for screening health services [ , ]. In the field of ophthalmology, there is now a significant burden of “catch-up” screening, critical for the early detection of blinding diseases before the onset of irreversible visual impairment (VI) [ ]. Glaucoma is one such leading cause of irreversible VI that is often detected late due to a lack of disease awareness and the often asymptomatic nature of early disease [ , ].Disruptions in patient care due to the pandemic necessitated the redesigning of clinical workflows to ensure continued clinical service delivery across many medical disciplines, including ophthalmology [
, , ]. This often required the incorporation of artificial intelligence, telehealth, and other related technologies across a spectrum of anatomical and functional eye screening processes [ ]. However, existing remote perimetry solutions have several limitations that may impede translation, such as the need for variable and costly specialized hardware, dependence on trained operators for testing, lack of robust comparative evidence, as well as technical limitations in detecting mild visual field deficits [ , ]. These limitations present challenges for the interpretation of results from current remote perimetry solutions and are likely to introduce significant barriers to adoption, particularly in community-based primary care settings [ ].In a recently published user acceptance test (UAT) we demonstrated patient acceptance of gamified software applied for ophthalmology-related health promotion and education in a mixed reality metaverse environment using augmented reality or virtual reality (VR) [
]. The software was designed for compatibility with commercially available VR headsets that are easily accessible to patients. We subsequently reconfigured the software for application in suprathreshold perimetry testing in a VR metaverse environment to meet test requirements such as fixed pupil-screen distance and control of background luminance. This software, Perispace (PS), was designed to meet the eye care needs of community-based primary care settings and general practitioner clinics, to perform remote perimetry tests and triage referrals of patients to tertiary ophthalmology eye screening services.In this proof-of-concept study, we performed a head-to-head comparison between the PS suprathreshold perimetry solution and the gold-standard Humphrey visual field (HVF) perimetry test.
Methods
Study Design
This was a prospective observational proof-of-concept study conducted at the National Health Care Group Eye Institute (NHGEI), Tan Tock Seng Hospital (TTSH), Singapore. Patients were included in this study only if they had a recent reliable HVF with tests performed within the prior 6 months. Criteria for reliable HVF test results in the manufacturer’s manual include less than 20% fixation losses, less than 33% false-negative errors, and less than 33% false-positive errors. Patients who were pregnant, unable to provide informed consent, or who had unreliable HVF perimetry results were excluded.
All eligible and willing participants received a comprehensive ophthalmic examination and a review of their recent gold standard HVF examination. HVF was performed using the 24-2 Swedish interactive threshold algorithm (SITA)–fast program [
] and visual field defects were classified as mild, moderate, or severe using the modified Hodapp, Parish, and Anderson criteria [ ]. Where patients performed HVF tests in both eyes, only the right eye was selected for analysis. PS perimetry tests were performed only after HVF tests.PS is a novel remote solution that replicates the photopic perimetry test of the HVF perimeter in a VR metaverse environment configured with clinical grade parameters including a background luminance of 10 cd/m2, age-adjusted stimulus luminance, and a stimulus angular size of 0.43°, corresponding to the Goldmann size III stimulus. Tests are performed by presenting stimuli to one eye at a time at random, to prevent malingering, with the VR headset calibrated using the Datacolor photometer. It has a wide, 120° field of view and uses PS software developed using the Unity engine (2020.3) to present test stimuli to one eye at a time at random.
Key Outcome Measures
The primary outcome measure was the agreement in overall classification of perimetry result (“normality” of result) between the HVF 24-2 (SITA-fast) and the Perispace v1.1 algorithm. Automated classification of perimetry results by the HVF has been extensively described and validated in earlier research [
]. Due to the pathophysiology of glaucoma, the disease has the propensity to affect arcuate fibers from the optic nerve head instead of the papillomacular (PM) bundle fibers or nasal radiating fibers [ ]. This is the basis for glaucomatous visual field defects—contiguous defects within the same hemifield with respect to the horizontal midline. Referral criteria applied to label an “abnormal” result based on topography in the PS perimetry results were based on the definition used in the landmark Ocular Hypertension Treatment Study (OHTS) [ ]. Therefore, all patients with >2 abnormal points detected in PS that were vertically or horizontally contiguous in a given hemifield, were labeled “abnormal” and referred for tertiary assessment.Secondary outcome measures for this study included the pointwise concordance of individual perimetry test points, visual comparison of perimetry topography maps, as well as the time taken for tests per eye, comparing results between PS and HVF perimetry tests. Pointwise concordance was calculated based on the agreement in automated classification of individual test points for all corresponding locations of the 24-2 test distribution, on both PS and HVF perimetry tests. Descriptive statistics are described, along with paired-sample 2-tailed t test to compare means of continuous variables. Statistical analysis was performed using SPSS (version 22; IBM Corp). P<.05 was considered statistically significant.
Ethical Considerations
The study was reviewed and approved by the Institutional Review Board of TTSH (reference nummber: IRB201800601), Singapore for de-identified data collection without financial compensation. Consecutive willing patients were recruited and informed consent was obtained.
Results
In this study, 10 consecutive subjects were recruited and analyzed. The average age of subjects was 59.6 (20.5) years (range 28-81 years). Of these subjects, 7 (70%) were male and 3 (30%) were female. Four (40%) subjects had pupil dilatation performed after the completion of HVF tests and before the conduct of PS tests. Demographics and clinical information are depicted in
. Based on gold standard HVF perimetry tests, 2 (20%) subjects were found to have no visual field defect, 4 (40%) subjects with mild visual field defect, 2 (20%) subjects with moderate visual field defect, and 2 (20%) subjects with severe visual field defect.Variable/category | Values | ||
Gender, n (%) | |||
Male | 7 (70) | ||
Female | 3 (30) | ||
Age, (years) mean (SD) | 59.6 (20.5) | ||
Age range (years) | 28-81 | ||
HVFa result, n (%) | |||
None | 2 (20) | ||
Mild | 4 (40) | ||
Moderate | 2 (20) | ||
Severe | 2 (20) | ||
Pupil dilation at the time of HVF and PSb test, n (%) | |||
None | 6 (60) | ||
Before both HVF and PS | 0 (0) | ||
After HVF, before PS | 4 (40) | ||
Lens status, N (%) | |||
Normal | 4 (40) | ||
Cataract | 4 (40) | ||
Intraocular lens | 2 (20) |
aHumphrey visual field.
bPerispace.
The agreement in overall classification of perimetry result normality between the HVF and PS was high (9/10, 90%). The pointwise concordance in the automated classification of individual test points was 83.3% (8.2%; range 75%-100%). The corresponding output perimetry greyscale topographic maps for HVF and that for PS had good topographic concordance for each of the 10 subjects, as depicted in
. The PS was reasonably able to delineate clinically significant visual field defects, such as superior and inferior arcuate scotomas (patients C and E), nasal step defects (patients B, I, and J), as well as central or paracentral defects (patients I and J), as illustrated in .The test time per eye was faster for PS than for HVF perimetry tests in 7 patients (7/10, 70%). The mean test time was 219.9 (29.4) seconds for the PS test, compared to 230.5 (29.8) seconds for the HVF. However, there was no statistically significant difference in mean test time (P=.25).
Discussion
Principal Findings
Clinical service disruptions arising as a result of the COVID-19 pandemic have exacerbated existing capacity shortages in community eye screening [
]. PS is a novel remote perimetry solution that replicates photopic perimetry calibrated with clinical grade configurations that address constraints with existing solutions. This study shows that remote perimetry using PS in a VR metaverse environment demonstrates good overall agreement, pointwise concordance, and topographic map concordance ( ) when compared with the gold standard HVF 24-2 SITA-fast perimetry test.These were critical outcome measures prioritized that were based on consensus from discussions with general practitioners and ophthalmic surgeons, to ensure out-of-hospital remote perimetry remains aligned with gold-standard perimetry and existing clinical referral workflows. Novel remote solutions that emerged in recent years have also demonstrated good performance when compared to the HVF [
]. Current leading solutions include the Eyecatcher in the United Kingdom, C3 Field Analyzer (CFA) in India, and VisuALL in the United States, among others [ - ]. Benefits of these perimetry solutions over the gold standard HVF include lower cost, portability, convenience, and patient acceptance [ , , ]. Existing remote perimetry tools can be broadly classified into VR-based and non–VR-based solutions.Limitations
The VisuALL and CFA tests are VR-based perimetry solutions using suprathreshold and threshold test strategies, respectively. Both tests exhibited good agreement in overall classification of perimetry result normality and concordance of perimetry topographic maps when compared against the HVF [
- ]. However, both the VisuALL and CFA tests appear to face limitations with regard to background and stimulus luminance. Both tools use background luminance in the scoptopic range, due to the limitations in the stimulus range of their hardware headset. This includes a background luminance of 3 cd/m2 reported for the VisuALL device and 4 cd/m2 reported for the CFA device, in contrast with the 10 cd/m2 of the gold standard HVF, which lies in the photopic range, approximating testing for daylight vision. The use of scotopic background luminance obfuscates the comparison of results with the HVF test and increases the risk of error due to the increased influence of absolute stimulus intensity on perception, being unable to account for pupil size and optic media effects [ ].The VisuALL is a promising solution that can be performed on the Pico VR headset with excellent performance, demonstrated previously [
, ]. However, the Pico headset only has a documented field of view of 100°, in comparison with the 108° required for the bilateral 24-2 HVF test distribution [ ]. Moreover, the adequacy of the stimulus range with the headset is unclear. It was reported as 120 cd/m2 in one trial [ ], but a later study reported a maximum intensity of 110 cd/m2 when plotted against RGB values [ ]. On the other hand, the CFA was occasionally unable to identify deficits that matched the HVF test potentially due to inaccurate stimulus luminance, given that investigators used the Luxmeter to perform calibration to 60 cd/m2 to approximate a fixed deficit of 18 dB, instead of using a photometer [ ].Non-VR headsets have other hardware limitations. The Eyecatcher perimetry solution uses a similar suprathreshold test strategy as the PS. However, it is unable to detect the 4 most central points of the 24-2 perimetry test distribution, due to the spatial imprecision of the sensors. This precludes testing in the crucial central-vision region, which can be impaired first in early glaucoma [
, ]. Of note, the Eyecatcher demonstrated good agreement with the HVF in overall classification of perimetry result normality, concordance of the individual test points with a mean of 83% (similar to the mean of 83.3% in PS), and concordance of perimetry topographic maps [ ]. However, the pointwise concordance with the Eyecatcher ranged from 50%-97%, compared to that for PS which was 75%-100%. This pointwise variation is similar to the spectrum of test-retest variability of HVF perimetry that has been reported on repeating the test due to contributing factors such as loss in patient attention, with worsening variability at the midrange of vision sensitivity [ ]. The authors of the Eyecatcher study also reported that hardware limitations such as limited spatiotemporal precision and imperfect gaze-calibration may contribute to the pointwise variation [ ]. Future research is needed to compare test-retest variability between perimetry devices and develop solutions to reduce variation such as by improving engagement using gamification and immersion in metaverse environments to maximize patient attention and effort with each run of the perimetry test.In addition to the above technical limitations, current solutions require skilled operators to reinforce testing instructions to ensure reliable results, as well as to operationalize requirements such as gaze calibration or eye patching for monocular testing. In addition, several non–VR-based solutions such as the Eyecatcher are tablet based and require additional specialized head- or eye-tracking hardware [
]. Moreover, unlike VR headsets, these do not have a closed cupola and have a restricted spatial range [ ]. This necessitates correction for eye movements, resulting in test-retest variations in the shape and extent of the testing and output topographic map [ ]. Therefore, they require scaling of testing stimuli to screen-size based on testing set-up (which may vary from one health care setting to the next), a dedicated room to prevent visual distractions, and adjustable lights dimmed for testing [ , ].Therefore, existing remote perimetry solutions described in the literature may not be suitable for primary care settings that are often manpower-, resource-, and space-constrained [
]. The strengths of the PS solution are that it addresses some of these limitations through automation of operator requirements such as monocular patching through algorithmic randomized differential testing of the patient’s eyes, calibration of testing parameters with the gold standard HVF using a photometer, as well as a closed cupola to provide sufficient spatial range, adequate ambient light control, and prevent external visual distractions. Moreover, with the automation of perimetry testing in a metaverse environment, this work has paved the way to decentralize the availability of crucial functional vision testing into out-of-hospital settings to overcome the limitations of physical space and infectious disease transmission risk in hospital settings [ ]. The limitations of this proof-of-concept study include a small cohort and a lack of control group for analysis. Despite these limitations, the results of this study justify further research using case-control methodology in larger, prospectively recruited cohorts stratified based on well patients or those with relevant comorbidities. These studies should be designed to examine the performance of remote perimetry solutions in discriminating patients requiring referral for further assessment or interventions, and standardized testing procedures (such as lack of pupil dilatation).Conclusions
In conclusion, remote perimetry using PS in a VR metaverse environment had good performance when compared against gold-standard HVF perimetry, based on overall agreement, pointwise concordance, and topographic map concordance. PS has several design benefits over existing remote perimetry solutions, that have been customized to avail perimetry for functional eye screening in out-of-hospital settings. This can help to facilitate the early detection of patients with VI in the community and prompt early referrals for formal evaluation in hospital settings to help address global trends in increasing preventable blindness [
].Acknowledgments
We acknowledge Pixel studios, Infocomm Media Development Authority Singapore, and Forbes Media LLC (30 under 30 Asia 2018 list team) for their kind support, which was critical to the success of this project. The study was partially funded by Enterprise Singapore and Tan Tock Seng Hospital, Singapore, which supported the research staffing hours contributed by study team members used by the hospital. Funders played no part in the design and conduct of the study, collection of the data, preparation or approval of the manuscript, nor the decision to submit the manuscript for publication.
Data Availability
The data set for this study is not publicly available due to local privacy regulations and a lack of patient consent for data sharing with public repositories.
Conflicts of Interest
Authors KAW, DVG, GSWT, and TYW report equity interest in the company VISRE (Visual Reality), which developed the Perispace version 1.1 solution. Author DVG also reports investment in DoctorBell (acquired by MaNaDr, Mobile Health), AskDr, Healthlink, and Shyfts. He declares receipt of travel funding from a Mobile Health Education grant, Commonwealth Fellowship in Innovation award, and National Youth Fund award for clinical research training and collaborations at Oxford University and Stanford University. He reports serving as senior lecturer and faculty advisor to the medical innovation program of the National University of Singapore, and appointment as Head (Digital Health) at Northeast Health International postinvestment by parent company Sheares Health Care. Author GSWT also reports clinical research grants from Santen and fees from Zeiss, Leica, Topcon, Optos-Nikon, Allergan, Roche, Novartis, and Bayer, governed by the policies of the Singapore National Eye Centre, as well as start-up equity in EyRiS. Author TYW also reports clinical trial grants from Allergan, Bayer, Boehringer-Ingelheim, Genentech, Merck, Novartis, Oxurion, Roche, and Samsung Bioepis; travel support from Allergan, Bayer, Boehringer-Ingelheim, Genentech, Merck, Novartis, Oxurion, Roche, and Samsung Bioepis; and personal fees from Allergan, Bayer, Boehringer-Ingelheim, Genentech, Merck, Novartis, Oxurion, Roche, and Samsung Bioepis, governed by the policies of the Singapore National Eye Centre; TYW is also a cofounder of the start-up companies Plano and EyRis and reports a patent for a method of modifying a retina fundus image for a deep learning model issued to EyRiS (WO2019022663).
References
- Foot B, MacEwen C. Surveillance of sight loss due to delay in ophthalmic treatment or review: frequency, cause and outcome. Eye (Lond). 2017;31(5):771-775. [FREE Full text] [CrossRef] [Medline]
- Teo ZL, Tham YC, Yu M, Chee ML, Rim TH, Cheung N, et al. Global prevalence of diabetic retinopathy and projection of burden through 2045: systematic review and meta-analysis. Ophthalmology. 2021;128(11):1580-1591. [FREE Full text] [CrossRef] [Medline]
- Low R, Lee JM, Lai SS, Rousselot A, Agarwal M, Agrawal R. Eye care during the COVID-19 pandemic: a report on patients' perceptions and experiences, an Asian perspective. Ophthalmol Ther. 2022;11(1):403-419. [FREE Full text] [CrossRef] [Medline]
- Kuehn BM. Despite improvements, COVID-19's health care disruptions persist. JAMA. 2021;325(23):2335. [CrossRef] [Medline]
- Whaley CM, Pera MF, Cantor J, Chang J, Velasco J, Hagg HK, et al. Changes in health services use among commercially insured US populations during the COVID-19 pandemic. JAMA Netw Open. 2020;3(11):e2024984. [FREE Full text] [CrossRef] [Medline]
- Arsenault C, Gage A, Kim MK, Kapoor NR, Akweongo P, Amponsah F, et al. COVID-19 and resilience of healthcare systems in ten countries. Nat Med. 2022;28(6):1314-1324. [FREE Full text] [CrossRef] [Medline]
- Broadway DC, Tibbenham K. Tackling the NHS glaucoma clinic backlog issue. Eye (Lond). 2019;33(11):1715-1721. [FREE Full text] [CrossRef] [Medline]
- Chua J, Baskaran M, Ong PG, Zheng Y, Wong TY, Aung T, et al. Prevalence, risk factors, and visual features of undiagnosed glaucoma: the Singapore epidemiology of eye diseases study. JAMA Ophthalmol. 2015;133(8):938-946. [FREE Full text] [CrossRef] [Medline]
- Vinod K, Sidoti PA. How glaucoma care changed for the better after the pandemic. Curr Opin Ophthalmol. 2022;33(2):59-66. [FREE Full text] [CrossRef] [Medline]
- Tham YC, Husain R, Teo KYC, Tan ACS, Chew ACY, Ting DS, et al. New digital models of care in ophthalmology, during and beyond the COVID-19 pandemic. Br J Ophthalmol. 2022;106(4):452-457. [FREE Full text] [CrossRef] [Medline]
- Gunasekeran DV, Tham YC, Ting DSW, Tan GSW, Wong TY. Digital health during COVID-19: lessons from operationalising new models of care in ophthalmology. Lancet Digit Health. 2021;3(2):e124-e134. [FREE Full text] [CrossRef] [Medline]
- Gunasekeran DV, Tseng RMWW, Tham YC, Wong TY. Applications of digital health for public health responses to COVID-19: a systematic scoping review of artificial intelligence, telehealth and related technologies. npj Digit Med. 2021;4(1):40. [FREE Full text] [CrossRef] [Medline]
- Iskander M, Ogunsola T, Ramachandran R, McGowan R, Al-Aswad LA. Virtual reality and augmented reality in ophthalmology: a contemporary prospective. Asia Pac J Ophthalmol (Phila). 2021;10(3):244-252. [FREE Full text] [CrossRef] [Medline]
- Ong CW, Tan MCJ, Lam M, Koh VTC. Applications of extended reality in ophthalmology: systematic review. J Med Internet Res. 2021;23(8):e24152. [FREE Full text] [CrossRef] [Medline]
- Gunasekeran DV, Low R, Gunasekeran R, Chan B, Ong HY, Raje D, et al. Population eye health education using augmented reality and virtual reality: scalable tools during and beyond COVID-19. BMJ Innov. 2021;7(2):278-283. [FREE Full text] [CrossRef] [Medline]
- Bengtsson B, Heijl A. SITA Fast, a new rapid perimetric threshold test. description of methods and evaluation in patients with manifest and suspect glaucoma. Acta Ophthalmol Scand. 1998;76(4):431-437. [FREE Full text] [CrossRef] [Medline]
- Hodapp E, Parrish RK, Anderson DR. Clinical Decisions in Glaucoma. Missouri, United States. Mosby; 1993.
- Hood DC, Raza AS, de Moraes CGV, Odel JG, Greenstein VC, Liebmann JM, et al. Initial arcuate defects within the central 10 degrees in glaucoma. Invest Ophthalmol Vis Sci. 2011;52(2):940-946. [FREE Full text] [CrossRef] [Medline]
- Keltner JL, Johnson CA, Cello KE, Edwards MA, Bandermann SE, Kass MA, et al. Ocular Hypertension Treatment Study Group. Classification of visual field abnormalities in the ocular hypertension treatment study. Arch Ophthalmol. 2003;121(5):643-650. [FREE Full text] [CrossRef] [Medline]
- Tan TF, Li Y, Lim JS, Gunasekeran DV, Teo ZL, Ng WY, et al. Metaverse and virtual health care in ophthalmology: opportunities and challenges. Asia Pac J Ophthalmol (Phila). 2022;11(3):237-246. [FREE Full text] [CrossRef] [Medline]
- Jones PR, Lindfield D, Crabb DP. Using an open-source tablet perimeter (Eyecatcher) as a rapid triage measure for glaucoma clinic waiting areas. Br J Ophthalmol. 2021;105(5):681-686. [FREE Full text] [CrossRef] [Medline]
- Razeghinejad R, Gonzalez-Garcia A, Myers JS, Katz LJ. Preliminary report on a novel virtual reality perimeter compared with standard automated perimetry. J Glaucoma. 2021;30(1):17-23. [CrossRef] [Medline]
- Montelongo M, Gonzalez A, Morgenstern F, Donahue SP, Groth SL. A virtual reality-based automated perimeter, device, and pilot study. Transl Vis Sci Technol. 2021;10(3):20. [FREE Full text] [CrossRef] [Medline]
- Mees L, Upadhyaya S, Kumar P, Kotawala S, Haran S, Rajasekar S, et al. Validation of a head-mounted virtual reality visual field screening device. J Glaucoma. 2020;29(2):86-91. [CrossRef] [Medline]
- Glen FC, Baker H, Crabb DP. A qualitative investigation into patients' views on visual field testing for glaucoma monitoring. BMJ Open. 2014;4(1):e003996. [FREE Full text] [CrossRef] [Medline]
- Heijl A, Patella VM, Bengtsson B. The Field Analyzer Primer: Excellent Perimetry, Fifth Edition. Jena, Germany. Carl Zeiss Meditec; 2021.
- Jones PR, Smith ND, Bi W, Crabb DP. Portable perimetry using eye-tracking on a tablet computer: a feasibility assessment. Transl Vis Sci Technol. 2019;8(1):17. [FREE Full text] [CrossRef] [Medline]
- Artes PH, Iwase A, Ohno Y, Kitazawa Y, Chauhan BC. Properties of perimetric threshold estimates from Full Threshold, SITA Standard, and SITA Fast strategies. Invest Ophthalmol Vis Sci. 2002;43(8):2654-2659. [Medline]
- Ratnarajan G, Newsom W, Vernon SA, Fenerty C, Henson D, Spencer F, et al. The effectiveness of schemes that refine referrals between primary and secondary care--the UK experience with glaucoma referrals: the Health Innovation and Education Cluster (HIEC) Glaucoma Pathways Project. BMJ Open. 2013;3(7):e002715. [FREE Full text] [CrossRef] [Medline]
Abbreviations
CFA: C3 field analyzer |
HVF: Humphrey visual field |
NHGEI: National Health Care Group Eye Institute |
NHI: Northeast Health International |
OHTS: Ocular Hypertension Treatment Study |
PM: papillomacular |
PS: Perispace |
SITA: Swedish interactive threshold algorithm |
TTSH: Tan Tock Seng Hospital |
VI: visual impairment |
VR: virtual reality |
Edited by T Leung; submitted 14.12.22; peer-reviewed by Y Chan, S Li; comments to author 20.03.23; revised version received 01.04.23; accepted 31.05.23; published 19.10.23.
Copyright©Kang-An Wong, Bryan Chin Hou Ang, Dinesh Visva Gunasekeran, Rahat Husain, Joewee Boon, Krishna Vikneson, Zyna Pei Qi Tan, Gavin Siew Wei Tan, Tien Yin Wong, Rupesh Agrawal. Originally published in the Journal of Medical Internet Research (https://www.jmir.org), 19.10.2023.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research, is properly cited. The complete bibliographic information, a link to the original publication on https://www.jmir.org/, as well as this copyright and license information must be included.