Abstract
Background: Epilepsy affects approximately 50 million people globally and imposes a substantial clinical and societal burden, requiring continuous and personalized monitoring for effective management. Wearable artificial intelligence (AI) technologies offer a promising solution by leveraging physiological signals and machine learning for seizure detection and prediction. While various approaches have been proposed, a comprehensive overview summarizing these advances and challenges is still needed.
Objective: This review aims to comprehensively explore and map the existing literature on AI-driven wearable technologies for epilepsy, identifying device characteristics, AI methodologies, biosignal measurements, validation approaches, and research gaps.
Methods: A scoping review was conducted following the PRISMA-ScR (Preferred Reporting Items for Systematic Reviews and Meta-Analyses Extension for Scoping Reviews) guidelines. A systematic search was performed across six electronic databases (Scopus, MEDLINE, Embase, ACM Digital Library, IEEE Xplore, and Google Scholar) to identify relevant studies published up to December 2023. We included studies that developed AI algorithms for epilepsy using noninvasive wearable devices (eg, smartwatches, smart clothing) and excluded those using nonwearables or in-body devices. Eligible publication types included journal articles, conference papers, and dissertations. Study selection and data extraction were performed independently by six reviewers. The extracted data were synthesized narratively.
Results: A total of 67 studies met the inclusion criteria. Research in this domain has increased significantly since 2021, with India, the United States, and China leading contributions. The studies examined both commercial (n=31, 46.3%) and noncommercial (n=31, 46.3%) wearable devices, with Empatica smart bands being the most frequently used. The primary biosignals monitored included activity measures (n=36, 53.7%), cardiovascular metrics (n=33, 49.3%), brain activity (n=24, 35.8%), and skin conductance (n=23, 34.3%). The most common AI models were support vector machines (n=28, 41.8%), random forests (n=14, 20.9%), and convolutional neural networks (n=10, 14.9%). Most models focused on seizure detection (n=54, 80.6%) compared to seizure prediction (n=14, 20.9%), reflecting a research imbalance that suggests the need for further development in predictive analytics. Sensitivity (n=54, 80.6%) was the most frequently reported performance metric, indicating a focus on identifying seizures; however, comprehensive clinical validation remains limited. Closed-source data predominated (n=44, 65.7%), limiting the generalizability of findings. The most used validation methods were leave-one-out cross-validation (n=21, 31.3%) and k-fold cross-validation (n=20, 29.9%), while video electroencephalography served as the primary reference standard (n=35, 52.2%).
Conclusions: Wearable AI technologies show significant promise in epilepsy management, offering real-time, continuous monitoring and early seizure detection. To realize clinical impact, future research should prioritize the standardization of validation methods, promote open data exchange for reproducibility, and develop energy-efficient algorithms that support real-world deployment in wearable devices.
doi:10.2196/73593
Keywords
Introduction
Background
Epilepsy affects an estimated 50 million individuals worldwide and remains one of the most common serious neurological disorders, often leading to significant physical injury, psychological distress, and reduced quality of life []. The condition’s prevalence exhibits geographical variation, with a higher incidence in low-income nations, often attributed to increased risk factors such as neurocysticercosis, perinatal complications, and limited health care access. Epilepsy encompasses various seizure types categorized broadly as focal, generalized, combined focal and generalized, or unknown onset epilepsy, each affecting patients differently [].
Accurate diagnosis and prompt management tailored to specific seizure types are crucial to minimizing risks and improving patient-related outcomes. Precise detection enables health care providers to develop tailored treatment strategies, potentially incorporating pharmacological interventions, lifestyle modifications, and, in certain cases, surgical procedures to achieve effective seizure control []. The current seizure detection techniques primarily focus on clinical assessments and electroencephalography (EEG). The former relies on patient histories and observed seizure events, while EEG records cerebral electrical activity to identify epilepsy-associated abnormalities. However, these methods have inherent limitations, such as limited accessibility, particularly in resource-constrained environments. EEG procedures are often time-intensive and expensive, and require patients to visit specialized centers. Moreover, the absence of seizure activity during EEG recording can yield inconclusive outcomes, underscoring the need for accessible and continuous monitoring solutions []. Given these limitations of conventional EEG, artificial intelligence (AI) emerges as particularly suitable for seizure detection due to its ability to analyze complex physiological data continuously and in real time. AI methods, such as machine learning algorithms, can automatically recognize subtle patterns within large volumes of physiological signals, enabling timely and accurate seizure detection and prediction. When integrated into wearable devices, AI offers a highly accessible and user-friendly approach that addresses the challenges of traditional EEG-based methods [].
Wearable AI-based devices represent a promising advancement in addressing the limitations of traditional epilepsy monitoring. These smart devices—such as smartwatches, wristbands, smart textiles, and head-mounted sensors—are designed to be worn on or near the body and enable continuous, noninvasive tracking of physiological data []. By leveraging real-time data collection and embedded AI algorithms, they can detect subtle changes in biosignals (eg, heart rate, movement, skin conductance) that may help to indicate the onset of a seizure. This can also facilitate early warnings and timely medical interventions, especially in unsupervised or at-home settings. Compared to conventional EEG, wearable AI solutions’ integration into daily life holds potential benefits to enhance patient autonomy, safety, and overall quality of life [].
This scoping review aims to map the current landscape of noninvasive, AI-driven wearable technologies for epilepsy detection and prediction, thereby providing a structured foundation for future investigations and facilitating translation of promising solutions into clinical practice.
Research Problem and Aim
Despite increasing interest in AI-based wearables for epilepsy, existing reviews remain fragmented and limited in scope. While several literature reviews have been conducted on this topic, they have been characterized by certain limitations. First, none of these reviews specifically targeted the intersection of AI and epilepsy detection [-]. Second, the reviews did not encompass data collected through wearable devices [,,,,]. Third, a significant portion of previous studies narrowly focused on a particular data type, predominantly EEG [,,]. Fourth, the majority of these studies were conventional literature reviews rather than more comprehensive scoping or systematic reviews [,,,,]. Lastly, the search strategies employed in these reviews often overlooked key databases such as CINAHL, PsycINFO, Scopus, IEEE, and the ACM digital library [,,,].
To address these gaps, the present review offers the first scoping review that systematically maps AI-driven, noninvasive wearable technologies used for epileptic seizure detection and prediction. The major contributions of this work include (1) mapping the landscape of wearable device types, biosignal modalities, and AI methodologies; (2) identifying validation frameworks and performance metrics; (3) identifying current research limitations and gaps; and (4) providing future directions, including clinical translation, regulatory challenges, and ethical considerations. This review aims to serve as a comprehensive and foundational resource to inform future research, development, and implementation of wearable AI systems in epilepsy management.
Methods
To achieve the objectives of this study, a scoping review was conducted following the guidelines of the PRISMA-ScR (Preferred Reporting Items for Systematic Reviews and Meta-Analyses Extension for Scoping Reviews) []; the PRISMA-ScR checklist associated with this review can be found in . The following sections provide a detailed description of the methods used in this review.
Search Strategy
To identify relevant studies, we performed searches across six electronic databases on December 7, 2023: Scopus, MEDLINE (via Ovid), Embase (via Ovid), ACM Digital Library, IEEE Xplore, and Google Scholar. These databases were selected to cover both clinical/biomedical (MEDLINE, Embase, and Scopus) and technical/engineering (ACM and IEEE) literature, ensuring broad coverage across disciplines relevant to AI and wearable technologies. We also set up a monthly automated search to run for 6 months, concluding on June 7, 2024. Given the large number of results generated by Google Scholar, which sorts entries by relevance, we limited our review to the top 100 entries (10 pages) to capture the most pertinent literature while maintaining feasibility and avoiding redundancy among different database searches. To ensure comprehensive coverage, we also performed backward and forward reference checking of included articles. Our search query consisted of three primary categories of terms: terms related to AI, terms related to epilepsy, and terms related to wearable devices. Our search query consisted of three key categories: terms related to AI, epilepsy, and wearable devices. Full search strategies for each database are included in .
Study Eligibility Criteria
This review considered studies that developed AI algorithms for epilepsy using data from noninvasive, body-worn wearable devices (eg, smartwatches, smart glasses, smart clothing, smart bracelets, smart tattoos). We excluded studies using nonwearables, handheld devices (eg, smartphones used independently), near-body wearables (eg, devices placed near but not on the body), and in-body wearables (eg, implants). Wired wearable systems were also excluded. We specifically included all AI algorithms used for any aspect of epilepsy, including diagnosis, monitoring, screening, therapy, prediction, and prevention. We included studies that combined data from wearable devices with other sources (eg, nonwearable devices, questionnaires, and interviews). We excluded studies that only presented a theoretical framework for AI-based wearable devices for epilepsy. Eligible publication types were peer-reviewed journal articles, conference papers, and dissertations written in English. We excluded reviews, preprints, conference abstracts, posters, protocols, editorials, and commentaries. No restrictions were imposed on measured outcomes, settings, publication dates, or countries of publication.
Study Selection Process
The study selection process was carried out in three stages. First, duplicates were removed from the retrieved studies using EndNote X9, followed by a manual duplicate check to ensure accuracy. Next, six reviewers independently screened titles and abstracts for relevance. Finally, the same six reviewers independently evaluated the full texts using predefined eligibility thresholds: the study had to involve AI methods, epilepsy-focused objectives, and data derived from wearable devices as per the criteria above. Discrepancies were resolved by a seventh reviewer.
Data Extraction Process
Six reviewers independently used Microsoft Excel to extract data on study metadata, wearable devices, epilepsy, and AI techniques. Any discrepancies between the reviewers were resolved through another reviewer by discussion. The extraction template included fields for study design, population, device characteristics, signal modalities, AI models, performance metrics, and clinical applications. The data extraction form can be found in .
Data Synthesis
Data extracted from the included studies were synthesized using a narrative approach, where information was summarized and presented through text, tables, and figures. To structure the synthesis, studies were categorized according to (1) the type of wearable device used (eg, wrist-worn, head-mounted), (2) biosignal modality (eg, EEG, photoplethysmography, electrodermal activity, accelerometer), (3) AI technique (eg, machine learning, deep learning), and (4) application context (eg, seizure detection, prediction, monitoring). The study began by outlining the metadata of the studies (eg, year and country of publication). Next, the characteristics of wearable devices used in the studies were outlined, including their status, type, placement, and operating system. Finally, the features of the AI techniques applied were detailed in terms of algorithms used, their objectives, dataset size, data input type, and model performance. Patterns and trends across categories were identified inductively through iterative comparison of extracted data.
Conflicting findings between studies (eg, differences in model accuracy or signal reliability) were documented and interpreted based on study context, population, data quality, and validation method. These contradictions were discussed in the Results and further reflected upon in the Discussion and Limitations sections. Data synthesis was managed using Microsoft Excel.
Results
Search Results
shows that when the above-mentioned databases were searched, 729 citations were returned. A total of 270 duplicates were identified and deleted using EndNote X9, leaving 459 studies. In addition, 87 studies were excluded after screening the titles and abstracts of these 459 studies. About 22 full-text studies were not able to be retrieved. After retrieving and reviewing the entire text of all 351 remaining studies, it was decided that 284 of them were ineligible for inclusion. The primary grounds for elimination were that they did not use wearable devices (n=198, 69.7%) or AI algorithms (n=24, 8.5%), were of unrelated publication genres (n=22, 7.7%), were not epilepsy-related research (n=36, 12.7%), or were nonhuman research (n=4, 1.4%). The study found no new studies related to this review by examining the reference list of the included studies. This review included a total of 67 studies [-].

Characteristics of Included Studies
provides an overview of the characteristics of the 67 included studies. Studies included in this analysis were published between 2013 and 2024. A high proportion of the articles were published in 2023 (n=16, 23.9%), followed by 2022 (n=14, 20.9%) and 2021 (n=13, 19.4%). In particular, 2016 and the early years have the smallest percentage of publications (n=4, 6.0%), which indicates an increasing interest in research in recent years. Geographically, India accounted for the largest number of studies (n=12, 17.9%), followed by the United States (n=9, 13.4%), China (n=8, 11.9%), Switzerland, Germany, Belgium, and Australia (each at n=5, 7.5%). According to the type of publication, most of the included studies were published as journal articles (n=40, 59.7%), while the remaining were conference papers (n=27, 40.3%). provides the characteristics of each included study.
| Features | Studies, n (%) | References |
| Year of publication | ||
| 2024 | 1 (1.5) | [] |
| 2023 | 16 (23.9) | [,,,,,,,-,,,-,] |
| 2022 | 14 (20.9) | [,,,,,,,,,,,,,] |
| 2021 | 13 (19.4) | [,,,,,,,,,,,,] |
| 2020 | 3 (4.5) | [,,] |
| 2019 | 7 (10.4) | [,,,,,,] |
| 2018 | 6 (9.0) | [,,,,,] |
| 2017 | 3 (4.5) | [,,] |
| ≤2016 | 4 (6.0) | [,,,] |
| Country of publication | ||
| India | 12 (17.9) | [,,,,,,,,,,,] |
| United States | 9 (13.4) | [,,,,,,,,] |
| China | 8 (11.9) | [,,,,,,,] |
| Switzerland | 5 (7.5) | [,,-] |
| Germany | 5 (7.5) | [-,,] |
| Belgium | 5 (7.5) | [,,,,] |
| Australia | 5 (7.5) | [,,,,] |
| Italy | 2 (3.0) | [,] |
| Sweden | 2 (3.0) | [,] |
| United Kingdom | 2 (3.0) | [,] |
| Other (n=1) | 12 (17.9) | [,,,-,,,,,,] |
| Publication type | ||
| Conference paper | 27 (40.3) | [-,,,-,,-,,,,,,,,,,,,,,] |
| Journal article | 40 (59.7) | [-,-,,,,-,,,-,-,,,,-,,-,,,,] |
aIncludes studies published in 2016 or earlier that add up to one study.
bIncludes studies from several countries that add up to one study.
Participant Demographics Across Included Studies
shows the demographic characteristics of the participants who were included in the study. The results indicate that 85.1% (n=57) of the articles reported the number of participants, which ranged from 1 to 400, with an average of 57 (SD 75.5) participants. The high SD value reflects a wide variation in sample sizes. Studies that did not report the number of participants were excluded from the average calculation (n=10, 14.9%). Regarding gender distribution, 40.3% (n=27) of the articles reported that the proportion of female participants ranged from 22% to 77%, with an average of 49.1% (SD 13.8%). However, 59.7% (n=40) of the studies did not provide the gender information, which presents a significant limitation in assessing the demographic balance. Additionally, only 26.9% (n=18) of the articles reported the age of the participants, with an average age of 25.4 (SD 12.7) years. The age range across studies extended from 1 month to 80 years, based on the lowest and highest age values reported. However, some of the articles did not report the participants’ exact age but mentioned participant age groups. The results of these articles showed that the age group of adult participants is most frequently included (n=31, 46.3%) because this category covers a wide age range from 19 to 64 years, followed by children (n=24, 35.8%) and older people (n=7, 10.4%), while 38.8% (n=26) do not specify the age groups of participants. provides the characteristics of each included study.
| Features | Studies, n (%) | Mean (SD) | Range | References |
| Number of participants | ||||
| Reported | 57 (85.1) | 57.0 (75.5) | 1‐400 | [-,-,-,,,,-,-,-] |
| Not reported | 10 (14.9) | — | — | [,-,,,,,,] |
| Age (y) | ||||
| Reported | 18 (26.9) | 25.4 (12.7) | 5.7‐45 | [,,,,,-,,,-,,,,,] |
| Not reported | 49 (73.1) | — | — | [-,,-,-,-,-,-,,,,,-,-,,,] |
| Female participants | ||||
| Reported | 27 (40.3) | 49.1 (13.8) | 22‐77 | [,,,,,,,-,,,,,-,,-,,-,] |
| Not reported | 40 (59.7) | — | — | [-,,,,,-,-,-,,,,,,,,-,-,,,] |
| Participants age groups | ||||
| Children | 24 (35.8) | — | 0‐18 | [,,,,,,,,,-,-,,,-,] |
| Adult | 31 (46.3) | — | 19‐64 | [-,,,,,-,,,,,,,-,,-,,,,] |
| Elderly | 7 (10.4) | — | ≥65 | [-,,,] |
| Not reported | 26 (38.8) | — | — | [-,,,,,-,-,,,,,,,,,,,] |
aAll means were calculated based on studies that reported the relevant data.
bNot applicable.
Characteristics of Wearable Devices
shows the characteristics of wearable devices, including their commercial status, device types, and placement. The result shows that the commercial (n=31, 46.3%) and noncommercial (n=31, 46.3%) wearable devices have equal percentages, with 7.5% (n=5) of studies not reporting the state of the device. Among the wearable device names, Empatica was the most commonly used (n=15, 22.4%), followed by Byteflies Kit (n=5, 7.5%), and 43.3% (n=29) did not report device names. Smart bands (n=24, 35.8%) were the most common type of wearable devices, followed by wearable sensors (n=18, 26.9%) and EEG devices (n=12, 17.9%). Regarding the placement of the wearable device, the result showed that half of the studies reported the devices being placed on the wrists (n=34, 50.7%), followed by the head (n=15, 22.4%), in the ear (n=10, 14.9%), and on the ankle (n=8, 11.9%). The features of wearable devices in each included study are presented in .
| Features | Studies, n (%) | References |
| Status of wearable device | ||
| Commercial | 31 (46.3) | [,,,-,,,,,,,-,,,,-,,,,,,,,-] |
| Noncommercial | 31 (46.3) | [,,-,-,-,,,,-,,-,,,,,,,,] |
| Not reported | 5 (7.5) | [,,,,] |
| Name of wearable device | ||
| Empatica | 15 (22.4) | [,,,,,,,,,,,,,,] |
| Byteflies Kit | 5 (7.5) | [,,,,] |
| Fitbit | 2 (3.0) | [,] |
| SmartCardia INYU | 2 (3.0) | [,] |
| Other (n=1) | 21 (31.3) | [,,,,,,,,-,,-,,] |
| Not reported | 29 (43.3) | [,-,-,,-,,,-,,,,,,,,,] |
| Type of wearable device | ||
| Smart band | 24 (35.8) | [-,,,,,,,,,,,,,,,-,] |
| Wearable sensor | 18 (26.9) | [,,,,,,,,,,,,,,-,] |
| EEG device | 12 (17.9) | [-,,-,,-] |
| Smartwatch | 7 (10.4) | [,,,,,,] |
| Other (n<3) | 6 (9.0) | [,,,,,] |
| Not reported | 3 (4.5) | [,,] |
| Placement of the wearable device | ||
| Wrist | 34 (50.7) | [-,,,,,,,-,,,,,-,,,-,,-,] |
| Head | 15 (22.4) | [,-,,,,,,,,,,] |
| Ear | 10 (14.9) | [,,,,,,,,,] |
| Ankle | 8 (11.9) | [,,-,,,] |
| Chest | 6 (9.0) | [,,,,,] |
| Arm | 5 (7.5) | [,,,,] |
| Back | 2 (3.0) | [,] |
| Other | 8 (11.9) | [,,,,] |
| Not reported | 7 (10.4) | [,,,,,,] |
aInclude different names of the wearable devices that add up to one study.
bEEG: electroencephalogram.
cEEG devices refer to EEG-based wearable technologies.
dInclude different types of wearable devices that add up to less than three studies.
shows the types of biosignals measured, sensors used, and sensing methods applied in the 67 included studies on wearable devices. The wearable devices of the included studies measured various biosignals. The most commonly measured biosignal was activity measurement (n=36, 53.7%), followed by cardiovascular measurement (n=33, 49.3%), brain activity (n=24, 35.8%), and skin conductance (n=23, 34.3%). On the other hand, the less commonly used biosignals include acoustic (n=3, 4.5%), orientation (n=3, 4.5%), respiratory measures (n=3, 4.5%), and sleep measures (n=1, 1.5%), which were less common. The wearable devices used various types of sensors, with accelerometers being the most frequent (n=41, 61.2%), followed by electroencephalogram sensors (n=24, 35.8%), electrodermal sensors (n=21, 31.3%), and photoplethysmography (n=18, 26.9%). The majority of studies (n=63, 94.0%) relied on an opportunistic approach for data sensing, with data automatically collected by devices without active user input and often collected through sensors embedded in mobile phones and wearables. A small number of studies (n=1, 1.5%) used a participatory approach, in which users actively contributed data by means of manual input, self-reporting, or signal reaction. The remaining 4.5% (n=3) of the studies did not clearly indicate the types of sensing methods used. The sensors of wearable devices included in each study are presented in .
| Features | Studies, n (%) | References |
| Measured biosignal | ||
| Activity measures | 36 (53.7) | [,-,,,,,,,,-,-,-,-,-,,,] |
| Cardiovascular measures | 33 (49.3) | [-,,,,-,,-,,,,,,,,-,-,,] |
| Brain activity | 24 (35.8) | [,,-,,,,,-,,,,,,,,-,] |
| Skin conductance | 23 (34.3) | [,,,,,,,,,,,,,,,-,-,,] |
| Skin temperature | 13 (19.4) | [,,,,,,,,,,,,] |
| Neuromuscular activity | 8 (11.9) | [,,,,,,,] |
| Acoustic | 3 (4.5) | [,,] |
| Orientation | 3 (4.5) | [,,] |
| Respiratory measures | 3 (4.5) | [,,] |
| Sleep measures | 1 (1.5) | [] |
| Not reported | 1 (1.5) | [] |
| Sensors | ||
| Accelerometer | 41 (61.2) | [,,-,,,,,,,-,-,-,-,-,-,] |
| Electroencephalogram | 24 (35.8) | [,,-,,,,-,,,,,,,,-,] |
| Electrodermal | 21 (31.3) | [,,,,,,,,,,,,,,,-,-,,] |
| Photoplethysmography | 18 (26.9) | [,,,,,,,,,,,,,,,,,] |
| Thermometer | 12 (17.9) | [,,,,,,,,,,,] |
| Electrocardiogram | 11 (16.4) | [,,,-,,,,,] |
| Gyroscope | 9 (13.4) | [,,,,,,,,] |
| Electromyogram | 6 (9.0) | [,,,,,] |
| Other (n<3) | 8 (11.9) | [,,,,,,,] |
| Sensing technology | ||
| Opportunistic | 63 (94.0) | [-,,-,,-,-] |
| Participatory | 1 (1.5) | [] |
| Not reported | 3 (4.5) | [,,] |
aBiosignals represent the physiological metrics measured by wearable devices. Sensor types refer to the specific technologies embedded in wearable devices.
bInclude different sensors of wearable devices that add up to less than three studies.
Features of AI
shows a problem-solving method, type of AI algorithm, and proposed function in all included studies (n=67). This finding indicates that classification was the main method used in all studies (100%). Support vector machines (SVMs; n=28, 41.8%) are the most widely used AI algorithms, followed by random forests (n=14, 20.9%), convolutional neural networks (CNNs; n=10, 14.9%), boosting models, and logistic regression (n=8, 11.9%). Less common methods include deep neural networks (n=7, 10.4%), K-nearest neighbors (n=6, 9%), repetition neural networks (n=9, 9%), naive Bayes (n=4, 6%), and multilayer perceptron (n=4, 6%). A small number of studies (n=5, 7.5%) used other algorithms, but 3% (n=2) did not report the algorithms used. As for the purpose of using AI, most studies focused on detecting epilepsy (n=54, 80.6%), while the target prediction rate was 22.4% (n=15; see ).
| Features | Studies, n (%) | References |
| Problem-solving approaches | ||
| Classification | 67 (100) | [-] |
| AI algorithms | ||
| Support vector machines | 28 (41.8) | [,,-,,,,,,,-,,,,,-,,,,,] |
| Random forest | 14 (20.9) | [,,,,,,,,,-,,] |
| Convolutional networks | 10 (14.9) | [,,,,,,,,,] |
| Boosting models | 8 (11.9) | [,,,,,,,] |
| Logistic regression | 8 (11.9) | [,,,,,,,] |
| Deep neural networks | 7 (10.4) | [,,,,,] |
| K-nearest neighbors | 6 (9.0) | [,,,,,] |
| Recurrent neural networks | 6 (9.0) | [,,,,,] |
| Naive Bayes | 4 (6.0) | [,,,] |
| Multilayer perceptron | 4 (6.0) | [,,,] |
| Other (n<3) | 5 (7.5) | [,,,,] |
| Not reported | 2 (3.0) | [,] |
| Aim of AI algorithms | ||
| Detection | 54 (80.6) | [,,-,-,,-,-,-,,-,,-,-] |
| Prediction | 15 (22.4) | [,,,,,,,,,,,,,,] |
aInclude different AI algorithms that add up to less than three studies.
summarizes the source and type of data, the nature of input signals transmitted into the AI model, and the number of features used in the included studies (n=67). Closed-source data was the most frequently used (n=44, 65.7%), and 28.4% (n=19) of studies used open-source data. The majority of the data (n=64, 95.5%) originated from wearable devices, while only 3% (n=2) was derived from non–wearable device data. In terms of data input to AI, the most common measurement is activity measures (n=34, 50.7%), followed by brain activity (n=24, 35.8%), electrodermal activity (n=21, 31.3%), and cardiovascular measurement (n=17, 25.4%). A few studies included other data inputs, such as electrocardiography (n=11, 16.4%), skin temperatures (n=10, 14.9%), electromyography (n=7, 10.4%), and blood oxygen saturation (n=2, 3.0%). While 1.5% (n=1) have not reported input types, 37.3% (n=25) of the articles have reported the number of features extracted from AI models that range from 2 to 1680, with an average of 111.7 (SD 329.4) features. presents the data characteristics and input features utilized in the AI models of each included study.
| Features | Studies, n (%) | References |
| Data source | ||
| Closed source | 44 (65.7) | [,,,,,,-,,,,,,-,,-,,,,-,-,] |
| Open source | 19 (28.4) | [-,,,,,,,,,,,,,-,] |
| Not reported | 4 (6.0) | [,,,] |
| Data types | ||
| WD-based data | 64 (95.5) | [-,,,-,-,-] |
| Non-WD–based data | 2 (3.0) | [,] |
| Not reported | 1 (1.5) | [] |
| Data input to AI algorithm | ||
| Activity measures | 34 (50.7) | [,,-,,,,,,,-,-,,,-,-,-,-] |
| Brain activity | 24 (35.8) | [,,-,,,,-,,,,,,,,-,] |
| Electrodermal activity | 21 (31.3) | [,,,,,,,,,,,,,,,,-,,] |
| Cardiovascular measures | 17 (25.4) | [,,,,,,,,,,,,,,,,] |
| Electrocardiography | 11 (16.4) | [,,-,,,,-] |
| Skin temperature | 10 (14.9) | [,,,,,,,,,] |
| Electromyography | 7 (10.4) | [,,,,,,] |
| Blood oxygen saturation | 2 (3.0) | [,] |
| Other (n=1) | 3 (4.5) | [,] |
| Not reported | 1 (1.5) | [] |
| Number of features | ||
| Reported | 25 (37.3) | [,,,,,,,,,,,,,,,,,,-,,,,] |
| Not reported | 42 (62.7) | [,-,,,,,-,,,,,,,,,-,-,-,-,,,,] |
aWD: wearable device.
bIncluded types of the data input to AI that add up to one study.
presents the reference standards, model validation methods, and performance metrics used in the included studies (n=67). The most commonly reported reference standard was the video recording with an EEG signal (n=35, 52.2%), followed by an EEG signal only (n=16, 23.9%). Only two studies (n=2, 3%) used electrocardiogram signals as a reference, and 20.9% (n=14) did not specify a reference standard. Different methods were used to evaluate and validate model performance. The leave-one-out cross-validation was the most common (n=21, 31.3%), followed by K-fold cross-validation (n=20, 29.9%) and training-test split (n=20, 29.9%). External validation has rarely been used (n=1, 1.5%), and 10.4% (n=7) of studies have not specified their validation method. Additionally, performance metrics were most commonly reported for sensitivity (n= 54, 80.6%) and accuracy (n=28, 41.8%). Additionally, metrics for specificity (n=24, 35.8%), false alarm rate (n=22, 32.8%), and precision (n=19, 28.4%) were frequently included. provides details on the AI features utilized in each of the cited studies.
| Featurs | Studies, n (%) | References |
| Reference standard | ||
| Video recording with EEG signals | 35 (52.2) | [,,,,,,,,,,-,,,,,,,,,-,,,,,-,,] |
| EEG signals | 16 (23.9) | [,,,,,,,,,,,,-,] |
| Electrocardiogram signals | 2 (3) | [,] |
| Not reported | 14 (20.9) | [,,,,,,,,,,,,,] |
| Model evaluation and validation | ||
| Leave-one-out cross-validation | 21 (31.3) | [,,,-,-,,,,,,-,,,,] |
| K-fold cross-validation | 20 (29.9) | [,,,,,,,,,,,-,,,,-] |
| Training-test split | 20 (29.9) | [,,,,,,,,,,,,,-,,,] |
| External validation | 1 (1.5) | [] |
| Not reported | 7 (10.4) | [,,,,,,] |
| Machine learning performance measures | ||
| Sensitivity | 54 (80.6) | [,,,-,-,-,-,-,-,-,,,] |
| Accuracy | 28 (41.8) | [,,-,,,,,,,,,,,,-,-,,,,] |
| Specificity | 24 (35.8) | [,,,,,-,,,,-,,,,,,,] |
| False alarm rate | 22 (32.8) | [,,-,,,,,,,-,,,,-] |
| Precision | 19 (28.4) | [,,,,,,-,,,,,,,,] |
| Area under curve | 10 (14.9) | [,,-,,,,,,] |
| False-positive rate | 7 (10.4) | [,,-,,] |
| F1-score | 11 (16.4) | [,,,,,,,,,,] |
| Negative predictive value | 3 (4.5) | [,,] |
| Detection latency | 3 (4.5) | [,,] |
| Other (n<3) | 8 (11.9) | [,,,,,,,,,] |
aEEG: electroencephalogram.
bVideo recording with EEG signals refers to the combined use of EEG and video monitoring for verifying seizure events.
cInclude different types of machine learning performance measures that add up to less than three studies.
Discussion
Principal Findings
This work analyzed 67 studies focused on wearable AI devices for epilepsy detection. A growing interest in using such devices from 2021 to 2023 was found, particularly in technologically advanced countries like India, the United States, and China. This trend underscores the increasing recognition of AI’s potential in epilepsy management. The comparable distribution of commercial (n=31, 46.3%) and noncommercial (n=31, 46.3%) wearable devices suggests developments from industry and research laboratories. Empatica and smart bands were dominant choices for activity and cardiovascular signal measurements. Also, the prevalent use of accelerometers and EEG sensors indicates their importance in improving the accuracy of seizure detection on the used devices. The use of various devices and biosignals highlights the versatility of wearable AI while also reflecting challenges such as data variability and device reliability.
The results indicated that SVMs (n=28, 41.8%), random forests (n=14, 20.9%), and CNNs (n=10, 14.9%) were the most commonly used AI models in wearable epilepsy research. SVMs were favored due to their robustness in handling high-dimensional physiological data with limited sample sizes, a common characteristic of wearable datasets. Random forests were used extensively due to their interpretability and ability to handle imbalanced datasets through ensemble learning. CNNs were widely used in research, taking advantage of time-series or image-like inputs from biosignals (ie, EEG or accelerometry), leveraging their spatial feature extraction. Such model selection is a trade-off between model complexity, interpretability, and ease of deployment on resource-constrained wearable devices. There were not many studies that utilized large-scale deep learning models, possibly due to the fact that large open datasets were scarce and energy efficiency was needed for wearable applications. It is found that studies employing ensemble methods (eg, boosting models: n=8, 11.9%) reported competitive performance. Such models could be further explored to enhance model detection accuracy.
The inclusion of multimodal data with diverse biosignals (eg, skin conductance: n=23, 34.3%; neuromuscular activity: n=8, 11.9%; respiratory measures: n=3, 4.5%) improved seizure detection rates. However, the scarcity of open-source datasets triggers the need for building comprehensive and accessible repositories. Few studies employed participatory sensing. Hence, there is an opportunity to incorporate patient feedback into data collection to enhance user engagement and personalization. Additional findings from our review reveal that smartwatches, though less common (n=7, 10.4%), have shown promising results in continuous seizure monitoring. On the other hand, EEG devices, despite being noninvasive, were often reported as less comfortable for long-term use compared to wrist-worn devices. Such a finding aligns with the need for user-centric design in medical wearables. Validation methods used for evaluating the AI models implemented in the wearable devices varied, with leave-one-out cross-validation (n=21, 31.3%) and K-fold cross-validation (n=20, 29.9%) being the most common ones. A frequent use of video EEG as a reference standard (n=35, 52.2%) was evident, which emphasizes the importance of reliable benchmarking in AI development. Most studies focused on adults (n=31%, 46.3%), with children and older participants being underrepresented. Performance metrics such as sensitivity (n=54, 80.6%), accuracy (n=28, 41.8%), and specificity (n=24, 35.8%) were widely reported, yet false alarms (n=22, 32.8%) remain a concern and indicate room for algorithmic refinement. Our review highlights critical gaps such as standardized validation, open data access, diverse demographic inclusion, and participatory sensing. Addressing these will enhance the clinical reliability and patient-centricity of AI-powered epilepsy management solutions.
Clinical Implications
The findings of this scoping review have significant clinical practice and real-world implications for the application of AI-based wearable technology in epilepsy treatment. Increased use of noninvasive wearable technology, particularly wrist-mounted devices such as smart bands and smartwatches, is an indication of more user-oriented continuous monitoring technology accommodated within patient lifestyles. This is a change that can improve compliance of patients, as these systems are less obtrusive and more comfortable than traditional EEG systems and enable long-term monitoring with minimal disruption to daily life. Real-time seizure detection, as demonstrated by the majority of studies reviewed (n=54, 80.6%), holds the potential for timely clinical intervention that might reduce the severity and damage of uncontrolled seizures. Early warning systems, although less established, would allow for patients and caregivers to institute preventive strategies in advance, such as rapid-acting medication or moving to a safe environment, and increase patient safety and autonomy. From an operations clinical perspective, wearable AI devices can decrease the need for prolonged hospital-based EEG monitoring, thereby cutting health care costs and burden. In addition, multimodal biosignal fusion (such as activity, cardiovascular, and electrodermal metrics) and interpretable AI models allow for the development of clinically explainable AI systems, which are a vital factor for doctor trust and uptake.
Research and Future Directions
This review underscores some research implications for future studies on wearable AI for epilepsy management. One such direction is related to the need to explore multimodal data collection that integrates biosignals (eg, cardiovascular measures, skin conductance, and neuromuscular activity) alongside brain activity measurements. Such an approach not only enhances the accuracy of seizure prediction models but also helps to develop early warning systems. Incorporating larger datasets from diverse demographics is also essential to improve the generalizability of the developed AI models.
Future research could focus on the development of user-friendly wearable devices (eg, smartwatches and wristbands) due to their higher user acceptance. Furthermore, devices capable of predicting seizures, besides the detection or monitoring tasks, will be particularly valuable. This could be achieved with wearable sensors equipped with advanced AI algorithms (ie, deep learning and large AI models). Providing timely alerts can allow patients and caregivers to take preventive measures that reduce seizure-related risks. One research direction could be focused on the development of large-scale deep learning frameworks that provide inference on the cloud without compromising device performance. Researchers could also focus on enhancing model transparency and interpretability to facilitate clinical adoption. Developing lightweight deep learning AI models, optimized for mobile and wearable devices, can provide real-time processing and large-scale data analysis that is essential for enhancing epilepsy management. Such models could be coupled with edge computing to minimize latency. One important implication is the need for more collaboration between AI developers, clinicians, and device manufacturers. Interdisciplinary approaches can ensure that AI algorithms are clinically relevant and can address the issues of false alarms and patient discomfort. Collaborative efforts can also lead to standardized protocols for data collection, validation, and sharing.
The ethical and regulatory aspects of AI in epilepsy care should be addressed. Future implications should explore frameworks for data privacy and security and algorithmic fairness. Integrating patient feedback into the design and implementation of wearable AI systems can also enhance their usability and acceptance. Security and privacy of the information are priorities due to the continuous collection and transmission of sensitive physiological data, often in real time. Wearables that record biosignals such as EEG, electrocardiogram, or electrodermal activity can unwittingly reveal aspects of a user’s neurological health, activity, and location. Lacking adequate encryption, secure storage centers in the cloud, and stringent access controls, these data streams may be vulnerable to breaches or misuse. They must comply with region-specific data privacy laws, such as the General Data Protection Regulation for the European region or the Health Insurance Portability and Accountability Act for the United States, and ensure that informed consent procedures are open and comprehensive. Regulator wise, the approval of AI-enabled wearables remains a moving target. The majority of existing frameworks, such as the Food and Drug Administration’s Digital Health Software Precertification Program or the European Medical Device Regulation, remain in catch-up mode with the complexity of dynamic, always-learning AI models. These kinds of systems oftentimes fall under Software as a Medical Device, which requires stringent validation for safety, efficacy, and risk management. Features for real-time seizure prediction or detection, especially those having therapeutic or alerting implications, will likely require follow-up clinical trials and postmarket surveillance to be approved by regulators.
Comparison to Prior Work
Compared to past reviews of seizure detection and wearable technology, this scoping review has a broader and more targeted synthesis at the intersection of AI, epilepsy, and noninvasive wearable technology. Past studies were prone to focus narrowly on EEG-based systems or novel machine learning techniques without considering the full gamut of wearable biosignals or device forms. For instance, the work by Li et al [] provided a general overview of wearable devices but did not rigorously review the AI algorithms or validation strategies implemented in epilepsy therapy. The work by Beniczky et al [] dealt with clinical practice guidelines without indicating the technical specifics of AI methods. Other works (eg, [,]) were primarily concerned with EEG-based seizure detection without regard to multimodal sensing or real-time deployment limitations. This work, on the other hand, brings together clinical and technical perspectives; includes both commercial and noncommercial devices; maps validation strategies, availability of datasets, and demographic consistency; and thus offers a more extensive and realistic foundation for the development of wearable AI for the treatment of epilepsy.
Strengths and Limitations
This review represents a systematic and methodologically rigorous synthesis of the wearable AI technology for the epilepsy field. Adhering to PRISMA-ScR guidance and systematically searching across six multidisciplinary databases, it includes a wide range of both engineering and clinical studies. Its coverage of more than one biosignal modality, AI algorithm, and device type represents an entire picture of the field. Moreover, the review highlights key research needs and provides unambiguous guidance for future development, such that it is a highly useful publication for researchers, clinicians, and device developers alike.
While this scoping review offers valuable insights into the landscape of wearable AI technologies for epilepsy, some limitations are to be considered when interpreting the findings. First, we only looked at wearable AI solutions in epilepsy, limiting generalizability to other neurological diseases that might be managed with similar technologies. Second, the majority of included studies were focused on detecting rather than predicting seizures. The resulting research imbalance limits us from forming strong conclusions about the performance of predictive models within the real-world environment. Third, our exclusion of in-body and near-body wearable devices, and nonwearable or handheld solutions, limits the review scope and possibly overlooks alternatives in epilepsy treatment. Fourth, we only looked at English articles after 2013, which may have led to overlooking prior or non–English-language studies on the issue. Fifth, most of the included studies used small sample sizes, closed-source data, and differing validation approaches. These render their findings less reproducible, weaker, and less clinically relevant. Lastly, by restricting the review to noninvasive wearables, we excluded studies using hybrid systems that may combine wearables with environmental sensors or mobile apps, which may offer a prolonged seizure management solution.
Conclusions
This review highlights the transformative potential of wearable AI technologies in epilepsy management, particularly in the detection and prediction of epileptic seizures. While machine learning algorithms like SVMs, random forests, and CNNs have shown notable performance, challenges remain regarding limited access to open-source datasets, inconsistent validation methods, and small sample sizes. Addressing these issues through open data sharing, standardized evaluation frameworks, and large-scale clinical validation will be critical.
The study emphasizes the importance of developing lightweight AI models optimized for real-time processing on wearable devices, while also exploring cloud-based solutions for complex analysis. Multimodal data integration—including physiological, behavioral, and contextual signals—can significantly enhance predictive performance and reliability, paving the way for proactive and personalized epilepsy care.
Future research should prioritize ethical considerations such as data privacy, informed consent, and algorithmic transparency, alongside regulatory compliance to ensure safe deployment in clinical settings. Cross-sector collaboration among AI developers, clinicians, engineers, and regulators will be essential to deliver trustworthy, user-centered wearable solutions that can meaningfully reduce the burden of epilepsy worldwide.
Acknowledgments
No external financial support or grants were received from any public, commercial, or not-for-profit entities for the research, authorship, or publication of this article.
Data Availability
All data generated during this study are provided in the multimedia appendices. This study did not involve the utilization of any custom code or mathematical algorithm.
Conflicts of Interest
None declared.
Search strategy.
DOCX File, 33 KBData extraction form.
DOCX File, 18 KBCharacteristics of each included study.
DOCX File, 85 KBFeatures of wearable devices and sensors.
DOCX File, 94 KBFeatures of artificial intelligence algorithms.
DOCX File, 114 KBFeatures of data used in artificial intelligence algorithms.
DOCX File, 79 KBPRISMA-ScR checklist.
DOCX File, 110 KBReferences
- Liu J, Zhang P, Zou Q, et al. Status of epilepsy in the tropics: an overlooked perspective. Epilepsia Open. Mar 2023;8(1):32-45. [CrossRef] [Medline]
- Fisher RS, Acevedo C, Arzimanoglou A, et al. ILAE official report: a practical clinical definition of epilepsy. Epilepsia. Apr 2014;55(4):475-482. [CrossRef] [Medline]
- Anwar H, Khan QU, Nadeem N, Pervaiz I, Ali M, Cheema FF. Epileptic seizures. Discoveries (Craiova). Jun 12, 2020;8(2):e110. [CrossRef] [Medline]
- Bazoukis G, Bollepalli SC, Chung CT, et al. Application of artificial intelligence in the diagnosis of sleep apnea. J Clin Sleep Med. Jul 1, 2023;19(7):1337-1363. [CrossRef] [Medline]
- Han K, Liu C, Friedman D. Artificial intelligence/machine learning for epilepsy and seizure diagnosis. Epilepsy Behav. Jun 2024;155:109736. [CrossRef] [Medline]
- Beniczky S, Wiebe S, Jeppesen J, et al. Automated seizure detection using wearable devices: a clinical practice guideline of the International League Against Epilepsy and the International Federation of Clinical Neurophysiology. Clin Neurophysiol. May 2021;132(5):1173-1184. [CrossRef] [Medline]
- Dash DP, Kolekar M, Chakraborty C, Khosravi MR. Review of machine and deep learning techniques in epileptic seizure detection using physiological signals and sentiment analysis. ACM Trans Asian Low-Resour Lang Inf Process. Jan 31, 2024;23(1):1-29. [CrossRef]
- Johansson D, Malmgren K, Alt Murphy M. Wearable sensors for clinical applications in epilepsy, Parkinson’s disease, and stroke: a mixed-methods systematic review. J Neurol. Aug 2018;265(8):1740-1752. [CrossRef] [Medline]
- Li W, Wang G, Lei X, Sheng D, Yu T, Wang G. Seizure detection based on wearable devices: a review of device, mechanism, and algorithm. Acta Neuro Scandinavica. Dec 2022;146(6):723-731. [CrossRef]
- Nair PP, Aghoram R, Khilari ML. Applications of artificial intelligence in epilepsy. Adv Neural Inf Process Syst. 2021;8(2):41-48. [CrossRef]
- Rukasha T, I Woolley S, Kyriacou T, Collins T. Evaluation of wearable electronics for epilepsy: a systematic review. Electronics (Basel). 2020;9(6):968. [CrossRef]
- Rasheed K, Qayyum A, Qadir J, et al. Machine learning for predicting epileptic seizures using EEG signals: a review. IEEE Rev Biomed Eng. 2021;14:139-155. [CrossRef] [Medline]
- Saminu S, Xu G, Zhang S, et al. Applications of artificial intelligence in automatic detection of epileptic seizures using EEG signals: a review. Artif Intelligence Applications. 2022;1(1):11-25. [CrossRef]
- Tricco AC, Lillie E, Zarin W, et al. PRISMA extension for scoping reviews (PRISMA-ScR): checklist and explanation. Ann Intern Med. Oct 2, 2018;169(7):467-473. [CrossRef] [Medline]
- Agrahri A, Tyagi A, Kumar D, Kusmakar S, Palaniswami M, Yan B. Detection of epileptic seizure using accelerometer time series data and hidden markov model. Presented at: 2022 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC); Jul 11-15, 2022:2426-2429; Glasgow, Scotland, United Kingdom. [CrossRef]
- Ahmed A, Ahmad W, Khan MJ, Siddiqui SA, Cheema HM. A wearable sensor based multi-criteria-decision-system for real-time seizure detection. Annu Int Conf IEEE Eng Med Biol Soc. Jul 2017;2017:2377-2380. [CrossRef] [Medline]
- Al-Bakri AF, Villamar MF, Haddix C, Bensalem-Owen M, Sunderam S. Noninvasive seizure prediction using autonomic measurements in patients with refractory epilepsy. Annu Int Conf IEEE Eng Med Biol Soc. Jul 2018;2018:2422-2425. [CrossRef] [Medline]
- Al-Hussaini I, Mitchell CS. SeizFt: interpretable machine learning for seizure detection using wearables. Bioengineering (Basel). Aug 2, 2023;10(8):918. [CrossRef] [Medline]
- Baghersalimi S, Teijeiro T, Aminifar A, Atienza D. Decentralized federated learning for epileptic seizures detection in low-power wearable systems. IEEE Trans on Mobile Comput. 2023;23(5):6392-6407. [CrossRef]
- Borujeny GT, Yazdi M, Keshavarz-Haddad A, Borujeny AR. Detection of epileptic seizure using wireless sensor networks. J Med Signals Sens. Apr 2013;3(2):63-68. [Medline]
- Böttcher S, Bruno E, Epitashvili N, et al. Intra- and inter-subject perspectives on the detection of focal onset motor seizures in epilepsy patients. Sensors (Basel). Apr 26, 2022;22(9):3318. [CrossRef] [Medline]
- Böttcher S, Bruno E, Manyakov NV, et al. Detecting tonic-clonic seizures in multimodal biosignal data from wearables: methodology design and validation. JMIR Mhealth Uhealth. Nov 19, 2021;9(11):e27674. [CrossRef] [Medline]
- Buettner R, Frick J, Rieg T. High-performance detection of epilepsy in seizure-free EEG recordings: a novel machine learning approach using very specific epileptic EEG sub-bands. Presented at: ICIS 2019 Proceedings: 40th International Conference on Information Systems; Dec 15-18, 2019; Munich, Germany. URL: https://aisel.aisnet.org/icis2019/is_health/is_health/16/ [Accessed 2025-10-22]
- Burelo K, Ramantani G, Indiveri G, Sarnthein J. A neuromorphic spiking neural network detects epileptic high frequency oscillations in the scalp EEG. Sci Rep. Feb 2, 2022;12(1):1798. [CrossRef] [Medline]
- Clarke S, Karoly PJ, Nurse E, et al. Computer-assisted EEG diagnostic review for idiopathic generalized epilepsy. Epilepsy Behav. Aug 2021;121(Pt B):106556. [CrossRef] [Medline]
- Coşgun E, Çelebi A. FPGA based real-time epileptic seizure prediction system. Biocybern Biomed Eng. Jan 2021;41(1):278-292. [CrossRef]
- Dhoot R, Gupta V, Jain R. Seizure detection mechanism in children. Presented at: 2023 International Conference on Applied Intelligence and Sustainable Computing (ICAISC); Jun 16-17, 2023:1-6; Dharwad, India. [CrossRef]
- Dong C, Ye T, Long X, et al. A two-layer ensemble method for detecting epileptic seizures using a self-annotation bracelet with motor sensors. IEEE Trans Instrum Meas. 2022;71:1-13. [CrossRef]
- dos Santos KR, de Abreu de Sousa MA, dos Santos SD, Pires R, Thome-Souza S. Differentiation between epileptic and psychogenic nonepileptic seizures in electroencephalogram using wavelets and support-vector machines. Appl Artif Intell. Dec 31, 2022;36(1). [CrossRef]
- Escobar Cruz N, Solarte J, Gonzalez-Vargas A. Automated epileptic seizure detection system based on a wearable prototype and cloud computing to assist people with epilepsy. In: Applied Computer Sciences in Engineering. Springer, Cham; 2018:204-213. [CrossRef]
- Fawzy M, Hussien A, Mostafa H. FPGA utilized implementation of epileptic seizure detection system based on wearable devices using dynamic partial reconfiguration. Presented at: 2022 10th International Japan-Africa Conference on Electronics, Communications, and Computations (JAC-ECC); Dec 19-20, 2022; Alexandria, Egypt. [CrossRef]
- Forooghifar F, Aminifar A, Atienza Alonso D. Self-aware wearable systems in epileptic seizure detection. Presented at: 2018 21st Euromicro Conference on Digital System Design (DSD); Aug 29-31, 2018:426-432; Prague. [CrossRef]
- Forooghifar F, Aminifar A, Atienza D. Resource-aware distributed epilepsy monitoring using self-awareness from edge to cloud. IEEE Trans Biomed Circuits Syst. Dec 2019;13(6):1338-1350. [CrossRef] [Medline]
- Forooghifar F, Aminifar A, Teijeiro T, et al. Self-aware anomaly-detection for epilepsy monitoring on low-power wearable electrocardiographic devices. Presented at: 2021 IEEE 3rd International Conference on Artificial Intelligence Circuits and Systems (AICAS); Jun 6-9, 2021:1-4; Washington, DC. [CrossRef]
- G G, Shukla R, Singh G, Sahani AK. A machine learning approach to the smartwatch-based epileptic seizure detection system. IETE J Res. Jan 2, 2024;70(1):791-803. [CrossRef]
- Ge Y, Hu D, Cui X, et al. Multimodal wearable device signal based epilepsy detection with multi-scale convolutional neural network. In: Cognitive Systems and Information Processing. Springer; 2024:70-80. [CrossRef]
- Glaba P, Latka M, Krause MJ, et al. EEG phase synchronization during absence seizures. Front Neuroinform. 2023;17:1169584. [CrossRef] [Medline]
- Gu Y, Cleeren E, Dan J, et al. Comparison between scalp EEG and behind-the-ear EEG for development of a wearable seizure detection system for patients with focal epilepsy. Sensors (Basel). Dec 23, 2017;18(1):29. [CrossRef] [Medline]
- Guo P, Yu H, Karicheri SG, et al. Empowering wearable seizure forecasting with scheduled sampling. Presented at: 2023 IEEE EMBS International Conference on Biomedical and Health Informatics (BHI); Oct 15-18, 2023; Pittsburgh, PA, USA. [CrossRef]
- Guo Y, Zhang Y, Mursalin M, Xu W, Lo B. Automated epileptic seizure detection by analyzing wearable EEG signals using extended correlation-based feature selection. Presented at: 2018 IEEE 15th International Conference on Wearable and Implantable Body Sensor Networks (BSN); Mar 4-7, 2018; Las Vegas, NV, USA. [CrossRef]
- Gupta S, Ranga V, Agrawal P. EpilNet: a novel approach to IoT based epileptic seizure prediction and diagnosis system using artificial intelligence. Adv Distributed Computing Artif Intelligence J. 2021;10(4):435-452. [CrossRef]
- B H, Rajarajeswari K, Sreeja GG, Nagarathna P. Brain tumor detection using machine learning techniques with internet of things. Presented at: 2022 IEEE 2nd Mysore Sub Section International Conference (MysuruCon); Oct 16-17, 2022:1-7; Mysuru, India. [CrossRef]
- Hamlin A, Kobylarz E, Lever JH, Taylor S, Ray L. Assessing the feasibility of detecting epileptic seizures using non-cerebral sensor data. Comput Biol Med. Mar 2021;130:104232. [CrossRef] [Medline]
- Hassan S, Mwangi E, Kihato PK. IoT based monitoring system for epileptic patients. Heliyon. Jun 2022;8(6):e09618. [CrossRef] [Medline]
- Heldberg BE, Kautz T, Leutheuser H, Hopfengartner R, Kasper BS, Eskofier BM. Using wearable sensors for semiology-independent seizure detection - towards ambulatory monitoring of epilepsy. Annu Int Conf IEEE Eng Med Biol Soc. Aug 2015;2015:5593-5596. [CrossRef] [Medline]
- Huang B, Abtahi A, Aminifar A. Lightweight machine learning for seizure detection on wearable devices. Presented at: ICASSP 2023 - 2023 IEEE International Conference on Acoustics, Speech and Signal Processing (ICASSP); Jun 4-10, 2023; Rhodes Island, Greece. [CrossRef]
- Jeyabharathi D, Kesavaraja D, Sasirekac D. iEpilepsy monitoring and alerting system using machine learning algorithm and WHMS. J Appl Res Technol. 2020;18(3):154-160. [CrossRef]
- Jiang P, Gao F, Liu S, et al. Longitudinally tracking personal physiomes for precision management of childhood epilepsy. PLOS Digit Health. Dec 2022;1(12):e0000161. [CrossRef] [Medline]
- Johansson D, Ohlsson F, Krýsl D, et al. Tonic-clonic seizure detection using accelerometry-based wearable sensors: a prospective, video-EEG controlled study. Seizure. Feb 2019;65:48-54. [CrossRef] [Medline]
- Khan M, Tom M, Yasser M, Basmaji T, Fraiwan L. Artificial intelligence guided wearable seizure detection device. Presented at: 2023 9th International Conference on Optimization and Applications (ICOA); Oct 5-6, 2023:1-5; Abu Dhabi, United Arab Emirates. [CrossRef]
- Kok XH, Imtiaz SA, Rodriguez-Villegas E. Assessing the feasibility of acoustic based seizure detection. IEEE Trans Biomed Eng. Jul 2022;69(7):2379-2389. [CrossRef] [Medline]
- Kueh SM, Kazmierski TJ. Low-power and low-cost dedicated bit-serial hardware neural network for epileptic seizure prediction system. IEEE J Transl Eng Health Med. 2018;6:2000209. [CrossRef] [Medline]
- Kusmakar S, Karmakar C, Yan B, Obrien T, Muthuganapathy R, Palaniswami M. Automated detection of convulsive seizures using a wearable accelerometer device. IEEE Trans Biomed Eng. Feb 2019;66(2):421-432. [CrossRef] [Medline]
- Kusmakar S, Karmakar CK, Yan B, O’Brien TJ, Muthuganapathy R, Palaniswami M. Onset detection of epileptic seizures from accelerometry signal. Presented at: 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); Jul 18-21, 2018; Honolulu, HI. [CrossRef]
- Mehta D, Deshmukh T, Sundaresan YB, Kumaresan P. Continuous monitoring and detection of epileptic seizures using wearable device. In: Tiwari S, Trivedi M, Mishra K, Misra A, Kumar K, editors. Smart Innovations in Communication and Computational Sciences Advances in Intelligent Systems and Computing, Vol 851. Springer:77-84. [CrossRef]
- Meisel C, El Atrache R, Jackson M, Schubach S, Ufongene C, Loddenkemper T. Machine learning from wristband sensor data for wearable, noninvasive seizure forecasting. Epilepsia. Dec 2020;61(12):2653-2666. [CrossRef] [Medline]
- Milosevic M, Van de Vel A, Bonroy B, et al. Automated detection of tonic-clonic seizures using 3-D accelerometry and surface electromyography in pediatric patients. IEEE J Biomed Health Inform. Sep 2016;20(5):1333-1341. [CrossRef] [Medline]
- Milosevic M, Van de Vel A, Bonroy B, et al. Detection of epileptic convulsions from accelerometry signals through machine learning approach. Presented at: 2014 IEEE 24th International Workshop on Machine Learning for Signal Processing (MLSP); Sep 21-24, 2014; Reims, France. [CrossRef]
- Mittlesteadt J, Bambach S, Dawes A, et al. Evaluation of an activity tracker to detect seizures using machine learning. J Child Neurol. Nov 2020;35(13):873-878. [CrossRef] [Medline]
- Motahar EB, Ishtiaque F, Wadud MSI. Accelerometer-based convulsive seizure detection using CNN. Presented at: 2021 IEEE International Conference on Biomedical Engineering, Computer and Information Technology for Health (BECITHCON); Dec 4-5, 2021:46-51; Dhaka, Bangladesh. [CrossRef]
- Munch Nielsen J, Zibrandtsen IC, Masulli P, Lykke Sørensen T, Andersen TS, Wesenberg Kjær T. Towards a wearable multi-modal seizure detection system in epilepsy: a pilot study. Clin Neurophysiol. Apr 2022;136:40-48. [CrossRef] [Medline]
- Nasseri M, Pal Attia T, Joseph B, et al. Ambulatory seizure forecasting with a wrist-worn device using long-short term memory deep learning. Sci Rep. Nov 9, 2021;11(1):21935. [CrossRef] [Medline]
- Onorati F, Regalia G, Caborni C, et al. Multicenter clinical assessment of improved wearable multimodal convulsive seizure detectors. Epilepsia. Nov 2017;58(11):1870-1879. [CrossRef] [Medline]
- Prathaban BP, Balasubramanian R, Kalpana R. A wearable foreSeiz headband for forecasting real-time epileptic seizures. IEEE Sensors J. 2021;21(23):26892-26901. [CrossRef]
- Qian L, Lu J, Li W, et al. MCU-enabled epileptic seizure detection system with compressed learning. IEEE Internet Things J. 2023;11(5):8771-8782. [CrossRef]
- R S, K A, Chellamani GK. Epilepsy detection using embedded machine learning. Presented at: 2023 9th International Conference on Advanced Computing and Communication Systems (ICACCS); Mar 17-18, 2023:1914-1922; Coimbatore, India. [CrossRef]
- Raj D, Rafeeque R, Jeevan A, Rhishitha T, Akshay S, Merry K. Towards accurate seizure detection: machine learning model for wearable technology. Presented at: 2023 9th International Conference on Smart Computing and Communications (ICSCC); Aug 17-19, 2023:622-626; Kochi, Kerala, India. [CrossRef]
- Regalia G, Onorati F, Lai M, Caborni C, Picard RW. Multimodal wrist-worn devices for seizure detection and advancing research: focus on the Empatica wristbands. Epilepsy Res. Jul 2019;153:79-82. [CrossRef] [Medline]
- Seethalakshmi V, Vijaya R, Pavithra J, Jency C, S SB, Jha P. Machine learning approaches for detecting epileptic seizures based on EEG signals. Presented at: 2022 IEEE 2nd Mysore Sub Section International Conference (MysuruCon); Oct 16-17, 2022:1-4; Mysuru, India. [CrossRef]
- Stirling RE, Grayden DB, D’Souza W, et al. Forecasting seizure likelihood with wearable technology. Front Neurol. 2021;12:704060. [CrossRef] [Medline]
- Sutanto E, Purwanto TW, Fahmi F, Yazid M, Shalannanda W, Aziz M. Implementation of closing eyes detection with ear sensor of Muse EEG headband using support vector machine learning. Int J Intelligent Eng Syst. 2023;16(1):460-473. [CrossRef]
- Tian F, Yang J, Zhao S, Sawan M. NeuroCARE: a generic neuromorphic edge computing framework for healthcare applications. Front Neurosci. 2023;17:1093865. [CrossRef] [Medline]
- Vandecasteele K, De Cooman T, Chatzichristos C, et al. The power of ECG in multimodal patient-specific seizure monitoring: added value to an EEG-based detector using limited channels. Epilepsia. Oct 2021;62(10):2333-2343. [CrossRef] [Medline]
- Varun SK, Reddy Bollu TK, Shrestha S. Automatic seizure detection employing machine learning-based late fusion techniques over behind-the-ear and the scalp EEG signals. Presented at: 2023 IEEE 4th Annual Flagship India Council International Subsections Conference (INDISCON); Aug 5-7, 2023:1-4; Mysore, India. [CrossRef]
- Vieluf S, Cantley S, Jackson M, Zhang B, Bosl WJ, Loddenkemper T. Development of a multivariable seizure likelihood assessment based on clinical information and short autonomic activity recordings for children with epilepsy. Pediatr Neurol. Nov 2023;148:118-127. [CrossRef] [Medline]
- Vieluf S, Hasija T, Kuschel M, Reinsberger C, Loddenkemper T. Developing a deep canonical correlation-based technique for seizure prediction. Expert Syst Appl. Dec 2023;234:120986. [CrossRef]
- Wang Q, Zhao H, Wang X, Sheng D, Sun B, Li S. Daily epileptic seizure detection algorithm based on multi-modal physiological signals. Presented at: 2022 5th International Conference on Communication Engineering and Technology (ICCET); Feb 25-27, 2022:100-106; Shanghai, China. [CrossRef]
- Xianji J, Yuyin Z, Weiming T, Zhongwei L. Recognition of epileptic seizures based on wristband wearable devices. Presented at: 2021 International Conference on Communications, Information System and Computer Engineering (CISCE); May 14-16, 2021:321-325; Beijing, China. [CrossRef]
- Yu S, El Atrache R, Tang J, et al. Artificial intelligence-enhanced epileptic seizure detection by wearables. Epilepsia. Dec 2023;64(12):3213-3226. [CrossRef] [Medline]
- Zhang J, Chatzichristos C, Vandecasteele K, et al. Automatic annotation correction for wearable EEG based epileptic seizure detection. J Neural Eng. Feb 28, 2022;19(1). [CrossRef] [Medline]
- Zsom A, Tsekhan S, Hamid T, et al. Ictal autonomic activity recorded via wearable-sensors plus machine learning can discriminate epileptic and psychogenic nonepileptic seizures. Annu Int Conf IEEE Eng Med Biol Soc. Jul 2019;2019:3502-3506. [CrossRef] [Medline]
Abbreviations
| AI: artificial intelligence |
| CNN: convolutional neural network |
| EEG: electroencephalography |
| PRISMA-ScR: Preferred Reporting Items for Systematic Reviews and Meta-Analyses Extension for Scoping Reviews |
| SVM: support vector machine |
Edited by Taiane de Azevedo Cardoso; submitted 07.Mar.2025; peer-reviewed by Abdullah Ayub, Chibuzo Onah, Daniel Filipe Borges; final revised version received 26.Jun.2025; accepted 30.Jul.2025; published 31.Oct.2025.
Copyright© Sarah Aziz, Amal A M Ali, Hania Aslam, Noor ul Ain, Amna Tariq, Zain Sohail, Sofia Murtaza, Hafiza Iqra Mahmood, Muhammad Irfan Wazeer, Fozia Murtaza, Alaa Abd-alrazaq, Mohammed Alsahli, Rafat Damseh, Rawan AlSaad, Tariq Shahzad, Arfan Ahmed, Javaid Sheikh. Originally published in the Journal of Medical Internet Research (https://www.jmir.org), 31.Oct.2025.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research (ISSN 1438-8871), is properly cited. The complete bibliographic information, a link to the original publication on https://www.jmir.org/, as well as this copyright and license information must be included.