Investigating the Cost-Effectiveness of Telemonitoring Patients With Cardiac Implantable Electronic Devices: Systematic Review

Investigating the Cost-Effectiveness of Telemonitoring Patients With Cardiac Implantable Electronic Devices: Systematic Review

Investigating the Cost-Effectiveness of Telemonitoring Patients With Cardiac Implantable Electronic Devices: Systematic Review

Review

1Department of Public Health and Primary Care, Ghent University, Gent, Belgium

2Department of Cardiovascular Sciences, Universiteit Leuven, Leuven, Belgium

3Department of Genetics, Pharmacology and Physiopathology of Heart, Blood Vessels and Skeleton (GENCOR), Antwerp University, Antwerp, Belgium

Corresponding Author:

Sarah Raes, MSc

Department of Public Health and Primary Care

Ghent University

Corneel Heymanslaan 10

Gent, 9000

Belgium

Phone: 32 9 332 83 59

Email: Sarah.Raes@UGent.be


Background: Telemonitoring patients with cardiac implantable electronic devices (CIEDs) can improve their care management. However, the results of cost-effectiveness studies are heterogeneous. Therefore, it is still a matter of debate whether telemonitoring is worth the investment.

Objective: This systematic review aims to investigate the cost-effectiveness of telemonitoring patients with CIEDs, focusing on its key drivers, and the impact of the varying perspectives.

Methods: A systematic review was performed in PubMed, Web of Science, Embase, and EconLit. The search was completed on July 7, 2022. Studies were included if they fulfilled the following criteria: patients had a CIED, comparison with standard care, and inclusion of health economic evaluations (eg, cost-effectiveness analyses and cost-utility analyses). Only complete and peer-reviewed studies were included, and no year limits were applied. The exclusion criteria included studies with partial economic evaluations, systematic reviews or reports, and studies without standard care as a control group. Besides general study characteristics, the following outcome measures were extracted: impact on total cost or income, cost or income drivers, cost or income drivers per patient, cost or income drivers as a percentage of the total cost impact, incremental cost-effectiveness ratios, or cost-utility ratios. Quality was assessed using the Consensus Health Economic Criteria checklist.

Results: Overall, 15 cost-effectiveness analyses were included. All studies were performed in Western countries, mainly Europe, and had primarily a male participant population. Of the 15 studies, 3 (20%) calculated the incremental cost-effectiveness ratio, 1 (7%) the cost-utility ratio, and 11 (73%) the health and cost impact of telemonitoring. In total, 73% (11/15) of the studies indicated that telemonitoring of patients with implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy ICDs was cost-effective and cost-saving, both from a health care and patient perspective. Cost-effectiveness results for telemonitoring of patients with pacemakers were inconclusive. The key drivers for cost reduction from a health care perspective were hospitalizations and scheduled in-office visits. Hospitalization costs were reduced by up to US $912 per patient per year. Scheduled in-office visits included up to 61% of the total cost reduction. Key drivers for cost reduction from a patient perspective were loss of income, cost for scheduled in-office visits and transport. Finally, of the 15 studies, 8 (52%) reported improved quality of life, with statistically significance in only 1 (13%) study (P=.03).

Conclusions: From a health care and patient perspective, telemonitoring of patients with an ICD or a cardiac resynchronization therapy ICD is a cost-effective and cost-saving alternative to standard care. Inconclusive results were found for patients with pacemakers. However, telemonitoring can lead to a decrease in providers’ income, mainly due to a lack of reimbursement. Introducing appropriate reimbursement could make telemonitoring sustainable for providers while still being cost-effective from a health care payer perspective.

Trial Registration: PROSPERO CRD42022322334; https://tinyurl.com/puunapdr

J Med Internet Res 2024;26:e47616

doi:10.2196/47616

Keywords



Background

The implantation rates of cardiac implantable electronic devices (CIEDs), including pacemakers and implantable cardioverter-defibrillators (ICDs), have increased over the last decades due to expanded indications and a progressively aging population [1]. To evaluate the clinical status of the patient and device functioning, current guidelines recommend that older patients with pacemakers should be evaluated every 3 to 12 months and patients with ICDs should be evaluated every 3 to 6 months [2]. This regimen imposes a considerable burden on patients and physicians if the patient is required to be seen in person.

Telemonitoring, referring to the process of using telecommunication and information technology to monitor the health status of a patient and device function from a distance, can reduce this burden by replacing some in-office visits with transmissions from the patients’ home [3]. Existing research indicated that telemonitoring is safe (eg, experiencing equal major adverse events to standard care) [4,5]. The advantages of telemonitoring include fewer inappropriate shocks for patients with ICDs [4,6] and fewer hospitalizations for patients with atrial arrhythmias and strokes [4,6,7]. Moreover, there is a rapid detection of cardiovascular events and device malfunction [5,7], leading to a time reduction between clinical decision and intervention [8].

Besides the effectiveness of telemonitoring, patient experience is essential in high-quality health care services. Overall, patients with pacemakers on telemonitoring reported positive experiences comparable to the experience of patients with in-hospital monitoring [9]. Telemonitored patients with pacemakers tended to receive less information about their diagnosis but no significant differences were found in other items, such as confidence in clinicians, treatment decision involvement, treatment satisfaction, and waiting time before admission [9]. Another study indicated that telemonitoring of patients with a cardiac resynchronization therapy defibrillator (CRT-D) was time-saving for both patients and physicians [10].

Cost-effectiveness analyses are important to quantify the value of new interventions, informing both medical decision-making and public policy [11]. However, cost-effectiveness analyses depend on the perspective considered. The different perspectives are the health care payer perspective (eg, Medicare or Medicaid and British National Health Service), the patient perspective, the provider perspective (eg, physician), and the society perspective. The health care payer and societal perspectives differ from each other as the societal perspective includes indirect nonmedical costs (eg, transport) [12].

Objectives

As cost-effectiveness analyses have shown heterogeneous results, it is still debatable whether telemonitoring is worth the investment relative to standard care. However, data on cost-effectiveness are important for health care payers to make decisions on the reimbursement of telemonitoring. Lack of reimbursement can be an important adoption barrier for new technology [13,14]. For these 2 reasons, this paper reviews the cost-effectiveness of telemonitoring, reviews how the results differ from different perspectives, and describes the key drivers of the cost-effectiveness of telemonitoring.


Overview

The review protocol was published by PROSPERO (International Prospective Register of Systematic Reviews; CRD42022322334). This systematic review was carried out in accordance with the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) reporting guideline of 2020 [15], and the PRISMA-ScR (Preferred Reporting Items for Systematic Reviews and Meta-Analyses extension for Scoping Reviews) [16], which can be found in the Multimedia Appendix 1. Guidelines for preparing a systematic review of health economic evaluations were followed [17].

Literature Search

For this review, PubMed, Embase, EconLit, and Web of Science Core Collection were systematically searched. The last search was performed on July 7, 2022. No filters (eg, publication date or type of study) were applied. Search strategies for all electronic databases can be found in Multimedia Appendix 2.

Search strings were developed based on explorations of databases and previous reviews. The following key concepts were translated into strings: (1) CIEDs, (2) telemonitoring, and (3) economic evaluations (eg, cost-effectiveness analyses and cost-utility analyses). The latter was based on a validated search filter, designed to identify economic evaluations, and was broadened for this study to maximize sensitivity [18]. The search terms for CIEDs and telemonitoring were based on existing reviews [19-21].

Study Selection

Studies were included if their primary focus was on the cost-effectiveness of telemonitoring patients with a CIED. The eligibility criteria were defined a priori for study selection (Textbox 1). The population, intervention, comparator, and outcome strategy was applied to describe the criteria. Only complete and peer-reviewed studies were included. Specific exclusion criteria included partial economic evaluations, systematic reviews or reports, and studies without standard care as a control group. Only studies published in English, Dutch, French, or German were eligible for inclusion. The reference lists of the included studies were searched manually to identify relevant studies. Two reviewers (SR and AP) independently screened the titles and abstracts of all records using Rayyan (Rayyan Systems Inc) [22]. After the initial screening, full texts were retrieved and screened for a second time. The second screening round was independently performed by 2 reviewers (SR and AP). Reasons for exclusion were documented (Figure 1). For both screening rounds, reviewers were blinded from each other’s decision, and disagreements were resolved through discussion.

Textbox 1. Eligibility criteria.

Inclusion criteria

  • Intervention
    • Cardiac implantable electronic devices: pacemaker, implantable cardioverter-defibrillator, cardiac resynchronization therapy defibrillator, cardiac resynchronization therapy pacemaker, and loop recorder
  • Comparator
    • Standard care
  • Study design
    • Complete health economic evaluations (within-trial and model-based)
  • Context
    • All settings
  • Language
    • English, French, German, or Dutch

Exclusion criteria

  • Intervention
    • Implantable pulmonary artery pressure monitor
  • Study design
    • Partial health economic evaluations (outcomes related to costs or effectiveness only)
  • Specific criteria
    • Systematic reviews, reports, commentaries, congress abstracts, protocols, and animal studies
Figure 1. Flowchart of the study selection.

Quality Assessment

Two researchers (SR and AP) independently evaluated the original papers using the Consensus Health Economic Criteria (CHEC) checklist to assess the risk of bias [23]. The CHEC checklist included 19 items. Any disagreement was resolved by discussion and consensus. Interpretation of the CHEC list can be found in Multimedia Appendix 3. The included studies were classified into 4 quality categories: excellent (score of 100%), good quality (score between 75% and 100%), moderate quality (score between 50% and 75%), and low quality (score <50%) [24].

Synthesis of Results

The study characteristics and main outcomes of the original papers are presented in the Results section. SR extracted all data. A data extraction sheet was developed using an existing template [17]. The following information was extracted from the included studies: study identification, general study characteristics, results, and authors’ conclusion. The principal outcome measures were health outcomes, cost or income outcomes (eg, the impact on total cost or income, cost or income drivers, cost or income drivers per patient, and cost or income drivers as a percentage of the total cost impact), and incremental cost-effectiveness ratios (ICERs) or cost-utility ratios.

To facilitate comparison across studies, the following adjustments and interpretations were made. First, the cost or income outcomes were presented per patient per year, and different currencies were converted to US Dollar (reference year: 2019 and reference country: United States) [25]. Second, perspectives were categorized into the health care payer perspective, patient perspective, provider perspective, and societal perspective. For the purpose of our study, the provider includes physicians who are directly involved in the care of patients with CIED.


Overview

The selection process is shown in Figure 1. From a total of 3305 publications, 15 (0.45%) unique publications were reviewed. Studies were excluded because one of the following reasons: (1) intervention: the paper did not describe telemonitoring patients with a CIED; (2) outcome: the paper contained only a cost analysis and not a cost-effectiveness analysis; and (3) study design or publication: the paper was a partial health economic evaluation, congress abstract, protocol, systematic review, animal study, or with no peer review.

Population

Characteristics of the included studies can be found in Table 1. All 15 (100%) studies had a primarily male population, except for the Nordland study, which had an almost equal sex distribution (Table 1) [26]. The mean age of the population with pacemakers was between 75 (SD 24.64) and 81 (SD 6.47) years. The mean age of patients with an ICD or CRT-D was between 61 (SD 12.6) and 69 (SD not calculated) years, except for the PREDICT RM study, where >50% of the population was aged >75 years [27]. Furthermore, of the 15 studies, 1 (7%) included only older patients (with a mean age of 81 years) with pacemakers [28], and 2 (13%) ICD or CRT-D studies only included patients with heart failure [11,29].

Table 1. Main characteristics of the included studies.
StudyAuthor and yearPatients, nPopulation characteristicsAge (years), mean (SD)Male participant (%)CIEDa type
Poniente [28]Bautista-Mesa et al [28], 202255Mean age of 81 years81 (6.47)69Pacemaker
PREDICT RM [27]Hummel et al [27], 201915,254N/Ab53% of participants aged ≥75 yearsc72ICDd
TARIFFe [30]Ricci et al [30], 2016209N/A69 (10.17)85ICD or CRT-Df
Nordland [26]Lopez-Villegas et al [26], 202050N/A74.8 (24.64)52Pacemaker
EVOLVOg [11]Zanaboni et al [11], 2013200Patients with heart failure66-69 (SD not reported)79ICD or CRT-D
MORE-CAREh [29]Boriani et al [29], 2016865Patients with heart failure66 (10)76CRT-D
Burri et al [31]Burri et al [31], 2013N/APatients with biventricular CRT-D65 (SD not reported)N/AICD or CRT-D
Raatikainen et al [32]Raatikainen et al [32], 200841N/A62 (10)83ICD
Al-Khatib et al [33]Al-Khatib et al [33], 2009151N/A63 (SD not reported)72ICD or CRT-D
CONNECTi [8]Crossley et al [8], 20111997N/A65 (12.1)71ICD or CRT-D
ECOSTj [34]Guédon-Moreau et al [34], 2014310N/A60.7 (12.6)90ICD
EuroEcok [13]Heidbuchel et al [13], 2015303Patients with new or replacement VVI-ICDl or DDD-ICDm62.4 (13.1)81ICD or CRT-D
SAVE-HMn trial [35]Perl et al [35], 2013115Patients with dual chamber pacemaker74 (9)57Pacemaker
SAVE-HMn trial [35]Perl et al [35], 201336Patients with ICD-implant due to primary prevention of sudden cardiac death62.5 (10)86ICD
Chew et al [36]Chew et al [36], 20201830N/A66 (SD not reported)88ICD or CRT-D
Dario et al [37]Dario et al [37], 20161171N/A77.5 (9)58Pacemaker
Dario et al [37]Dario et al [37], 2016930N/A67.5 (12)79ICD

aCIED: cardiac implantable electronic device.

bN/A: not applicable.

cAge was a discrete variable in this study (higher of lower than 75 years old).

dICD: implantable cardioverter-defibrillator.

eTARIFF: Health Economics Evaluation Registry for Remote Follow-Up.

fCRT-D: cardiac resynchronization therapy defibrillator.

gEVOLVO: Evolution of Management Strategies of Heart Failure Patients With Implantable Defibrillators.

hMORE-CARE: Monitoring Resynchronization Devices and Cardiac Patients.

iCONNECT: Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision.

jECOST: Effectiveness and Cost of ICD Follow-Up Schedule With Telecardiology.

kEuroEco: European Health Economic Trial on Home Monitoring in ICD Patients.

lVVI-ICD: single-chamber ICD.

mDDD-ICD: dual-chamber ICD.

nSAVE-HM: Socio-Economic Effects and Cost Saving Potential of Remote Patient Monitoring.

Study Designs

Tables 2 and 3 show the summary table of results. Of 15 studies, 11 (73%) were conducted in Europe [11,26,28-32,34,35,37], 3 (20%) in the United States [27,33,38], and 1 (7%) in Canada [36]. Of the 15 studies, 3 (20%) calculated the ICER [26-28], 1 (7%) calculated the cost-utility ratio [11], and 11 (73%) calculated the cost impact of telemonitoring. All studies analyzed the health care payer perspective, with 33% (5/15) analyzing the patient perspective [11,28,30,32,34], 13% (2/15) analyzing the societal perspective [33,35], and 13% (2/15) analyzing the provider perspective [13,30].

Table 2. Summary of the main results.
StudyCountryDesignTime horizonCIEDa typeEffectCost-effectiveness in original currency and in reference year (in US $, 2019)Conclusion
Pacemaker studies

Poniente [28]SpainNon-RCTb5 yearsPacemakerQALYc difference: 0.27ICERd: €301.16 per QALY (US $270.09 per QALY)Cost-effective

Nordland [26]NorwayRCT1 yearPacemakerQALY difference: 0.03ICER: €53,345 per QALY (US $59.746 per QALY)Not cost-effective

SAVE-HMe trial [35]AustriaRCT17 monthsPacemakerNo adverse effects differenceN/AfCost-saving

Dario et al [37], 2016ItalyNon-RCT1 yearPacemakerAverage time reduction to treat patients (−4.1 minutes/follow-up)N/ACost-saving
ICDg or CRT-Dh studies

PREDICT RM [27]United StatesReal-worldLifelongICDQALY difference: 0.64ICER: US $10,752 per QALY (US $12,069 per QALY)Cost-effective

TARIFFi [30]ItalyNon-RCT12 monthsICD or CRT-DQALY difference: 0.02Not calculated because QALY difference was not significant (P=.53)Cost-saving

EVOLVOj [11]ItalyRCT16 monthsICD or CRT-DQALY difference: 0.066k (P=.03)Cost-utility ratio <0Dominant

MORE-CAREl [29]Europe and IsraelRCT2 yearsCRT-DQOLm difference:−1N/ACost-saving

Burri et al [31], 2013United KingdomSystematic review data10 yearsICD or CRT-DInappropriate shocks: −51%; battery exhaustion: −7%N/ACost-saving

Raatikainen et al [32], 2008FinlandNon-RCT18 monthsICDTime burden for patients of −175 minutes and physician of −17 minutes/patient/follow-upN/ACost-effective

Al-Khatib et al [33], 2009United StatesRCT1 yearICD or CRT-DEuroQoL difference: 25%; no difference in satisfaction and mortalityN/ACost-saving

CONNECTn [8]United StatesRCT15 monthsICD or CRT-DTime from clinical event to clinical decision: 17.4 daysk (P<.001)N/ACost-saving

ECOSTo [34]FranceRCT27 monthsICDPhysical, psychological, and SF-36p QOL scores: not significantN/ACost-saving

EuroEcoq [13]Belgium, Finland, Germany, United Kingdom, Spain, and the NetherlandsRCT2 yearsICD or CRT-DSF-36 QOL score: not significantN/ACost-saving

SAVE-HM trial [35]AustriaRCT26 monthsICDNo adverse effects differenceN/ACost-saving

Chew et al [36], 2022CanadaNon-RCT5 yearsICD or CRT-DRisk of death (hazard ratio): 0.43k (P<.001)N/ACost-saving

Dario et al [37], 2016ItalyNon-RCT1 yearICDAverage time reduction to treat patient (−13.7 minute/follow-up)N/ACost-saving

aCIED: cardiac implantable electronic device.

bRCT: randomized controlled trial.

cQALY: quality-adjusted life year.

dICER: incremental cost-effectiveness ratio.

eSAVE-HM: Socio-Economic Effects and Cost Saving Potential of Remote Patient Monitoring.

fN/A: not applicable.

gICD: implantable cardioverter-defibrillator.

hCRT-D: cardiac resynchronization therapy defibrillator.

iTARIFF: Health Economics Evaluation Registry for Remote Follow-Up.

jEVOLVO: Evolution of Management Strategies of Heart Failure Patients With Implantable Defibrillators.

kThe values are statistically significant.

lMORE-CARE: Monitoring Resynchronization Devices and Cardiac Patients.

mQOL: quality of life.

nCONNECT: Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision.

oECOST: Effectiveness and Cost of ICD Follow-Up Schedule With Telecardiology.

pSF-36: The 36-Item Short Form Survey.

qEuroEco: European Health Economic Trial on Home Monitoring in ICD Patients.

Table 3. Summary table of results related to perspectives and key cost or income drivers.
Study and perspectiveTotal cost or incomea impact compared to standard care in original currency in reference year (US $ ppb per year, 2019) and cost or incomea impact driversCost or incomea impact drivers pp (US $ pp per year, 2019)Cost or incomea impact drivers as a percentage of total cost or income impact (%)
Pacemaker studies

Poniente [28]


Health care payer perspective



, 2012: −€8 (−US $8.96)




Staff costs−€3.7 (−US $ 4.56)49




Ambulance transport−€3.2 (−US $3.9)42




Consultation room−€0.8c (−US $0.9)10


Patientperspective



, 2012: −€9 (−US $11.2)




Informal transport−€5.1c (−US $ 6.2)58




Lost income−€3.7 (−US $4.58)42

Nordland [26]


Health care payer perspective



, 2015: €1,808 (US $2,183)




Hospitalization€1,808.31 (US $2,183)100




Ambulance transport−€60 (−US $72.5)−3




Physician cost€39.39c(US $47.6)2




Consultation room€20.17c (US $24.3)1

SAVE-HMd trial [35]


Societalperspective




, 2013: −€914 (−US $1,113)





Transport−€911.3 (−US $1,020)99.7





Follow-up personnel cost−€26.7 (−US $32.93)2

Dario et al [37], 2016


Health care payer perspective



€, 2011: −€832c (−US $1,054)




Acute hospitalization−€816c (−US $1,034)98




Pharmacy medication−€26 (−US $32.93)3




EDe admission−€11.89 (−US $15.01)1




Visits and procedure€22.29 (US $28.22)−3
ICDf or CRT-Dg studies

PREDICT RM [27]


Health care payer perspective



US $, 2006: −$566 (−US $635)




Hospitalization−US $554 (−US $621.94)98




Nonhospital cost−US $12 (−US $13.44)2

TARIFFh [30]


Health care payer perspective



, 2011: −€562 (−US $712)




Cardiovascular hospitalization−€454c (−US $575.1)80




Scheduled visit, protocol based−€64.24c (−US $81.4)11




Outpatient diagnostic test−€36.93c (−US $46.82)7




Unscheduled visit€12.27c (US $15.6)−2




Emergency visit costs−€15.67 (−US $19.8)0.03




Outpatient clinical evaluation−€3.12 (−US $3.9)0.005


Patientperspective



, 2011: −€68 (−US $86)




Patient loss of work−€42.34c (−US $53.6)62




Traveling−€25.86c (−US $32.8)38


Providerperspective



, 2011: −€55 (−US $69)




Scheduled visit, protocol based−€64.24c (−US $81.4)117




Unscheduled visit€12.27c (US $15.6)−22




Outpatient clinical evaluation−€3.12 (−US $3.9)6

EVOLVOi [11]


Health care payer perspective



, 2010: −€167 (−US $219.5)




Hospitalization−€223 (−US $292.5)134




Scheduled visit, protocol based−€33.66c (−US $44.1)20




EDh and urgent visit−€8.81c (−US $11.5)5




Nonurgent in-office visit€10.68 (−US $14)−6




Diagnostic examinations−€0.56 (−US $0.78)0




Scheduled remote visit€32.20c (−US $42.2)−19




Unscheduled remote visit€56.42c (−US $74)−34


Patientperspective



, 2010: −€90 (−US $117)




Scheduled visit, protocol based−€96.90c (−US $127.6)110




ED and urgent visit−€23.81c (−US $31.2)27




Nonurgent visit€30.74 (US $40.32)−35

MORE-CAREj [29]


Health care payer perspective



, 2014, no reimbursement: −€62.5 (−US $76)




Cardiovascular hospitalization−€44.3 (−US $53.76)71




Scheduled visit, protocol based−€37.4 (−US $45.4)61




ED visits−€0.5 (−US $0.56)−1




Unscheduled visit€6.4 (US $7.8)−10




Device hospitalization€13.3 (US $16.1)−11



, 2014, with reimbursement: −€44.3 (−US $18)




Cardiovascular hospitalization−€44.3 (−US $53.8)306




Unscheduled remote checkMaximum −€29.4 (−US $35.7)203




Scheduled remote checkMaximum −€18.6 (−US $22.5)128




Scheduled visit, protocol based−€37.4 (−US $45.4)39




ED visits−€0.5 (−US $0.6)−1




Unscheduled visit€6.4 (US $7.8)−44




Device hospitalization€13.3 (US $16.1)−93

Burri et al [31], 2013


Health care payer perspective



£, 2007:−£3.3 (−US $6.7)




Initial investment and in-hospital follow-up visit−£3.3 (−US $6.7)100

Raatikainen et al [32], 2008


Health care payer perspective



, 2006: −€641 (−US $914)




In-office visitk, only 1 visit is protocol based−€560.0 (−US $798.1)87




Travelingk−€198.7 (−US $283.1)31




Remote monitoringk€146.7 (−US $208.9)−23




Accommodationk−€1.3 (−US $1.9)0




Sickness allowancek−€28.0 (−US $39.9)4


Patientperspective



, 2006:−€59 (−US $84)




Patient feek−€58.7 (−US $83.6)100

Al-Khatib et al [33], 2009


Societalperspective



US $, 2009: −US $254 (−US $245)




Patient loss of work−US $383 (−US $370.3)150




Traveling−US $19 (−US $18.4)7




Follow-up visit, only 1 visit is protocol basedUS $148 (US $ 143)−58

CONNECTl [8]


Health care payer perspective



US$, 2008: −US $1,434 (−US $1,243)




Mean cost per hospitalization−US $1,434.4 (−US $1,243)100

ECOSTm [34]


Health care payer perspective



, 2011: −€927 (−US $1,175)




Cardiovascular hospitalization−€720 (−US $912.3)78




Device cost−€533 (−US $675)58




Nonhospital cost−€227c (−US $287.6)24




Other nonhospital cost−€182 (−US $230.6)20




Cardiovascular treatment−€113 (−US $143.1)12




Device management cost−€74c (−US $93.7)8




ICD ambulatory visit, 3 visits are protocol based−€40c (−US $50.6)4




Traveling−€50 (−US $63.4)0.05




Reimbursement€1000 (US $1,266)−108


Patientperspective



, 2011: −€9 (−US $11.2)




Traveling−€9 (−US $11.4)100

EuroEcon [13]


Providerperspective



, 2013: −€0.5 (−US $1.12)




Follow-up visit, protocol based−€0.5 (−US $0.56)100


Health care payer perspective



, 2013: −€287 (−US $349)




Hospitalization−€301 (€326.9)105




Follow-up visit, protocol based−€0.5 (−US $0.56)−0.1




Examination€1.5 (US $1.8)−0.5




Other physician visit€12.5 (US $15.2)−4

SAVE-HM triald [35]


Societalperspective



€, 2013: −€804.9c (−US $981)




Transport−€787 (−US $959)98




Follow-up personnel cost, protocol based−€17 (−US $21.3)2

Chew et al[36], 2022


Health care payer perspective



CAD $, 2019: −$709.6 (−US $535.4)




Hospitalization−$682.4c (−US $514.4)96




EDe visits−$10.2 (−US $7.68)1




In-office visits−$17 (−US $12.8)2

Darioet al [37], 2016


Health care payer perspective



, 2011: −€338 (−US $428)




Hospitalization−€295 (−US $374)96




Visits and procedure, protocol based−€40 (−US $50.6)12




Pharmacy medication−€31 (−US $39.3)9




Hospital medication€1.37c (US $1.8)−0.45




EDe admission€27c (US $34.2)−8

aIf the perspective is health care system or patient, then cost and if the perspective is provider, then income.

bpp: per patient (in health care and patient perspectives) or per physician (in provider perspective).

cThe values are statistically significant.

dSAVE-HM: Socio-Economic Effects and Cost Saving Potential of Remote Patient Monitoring.

eED: emergency department.

fICD: implantable cardioverter-defibrillator.

gCRT-D: cardiac resynchronization therapy defibrillator.

hTARIFF: Health Economics Evaluation Registry for Remote Follow-Up.

iEVOLVO: Evolution of Management Strategies of Heart Failure Patients With Implantable Defibrillators.

jMORE-CARE: Monitoring Resynchronization Devices and Cardiac Patients.

kCosts were recalculated per patient.

lCONNECT: Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision.

mECOST: Effectiveness and Cost of ICD Follow-Up Schedule With Telecardiology.

nEuroEco: European Health Economic Trial on Home Monitoring in ICD Patients.

Intervention and Comparator

Telemonitoring entailed data transmission and data review. Table 4 shows the frequencies of data transmission, review, and in-office visits of the included studies. In 47% (7/15) of the studies, data were transmitted continuously or daily [26,28,30,31,34,35]; in 20% (3/15) studies, data were transmitted after a device alert [8,11,29]; and in 13% (2/15) studies, data were transmitted every 3 months [32,33]. In 20% (3/15) of the studies, data review was performed daily [28,34,35]; however, in 40% (6/15) of the studies, it was performed after a device alert was received [8,11,26,29-31]. Besides data transmission and review, telemonitoring included scheduled in-office visits. In 33% (5/15) of the studies, all scheduled in-office visits were based on the protocol [11,13,29,30,35]. In 7% (1/15) of the studies, at least 1 scheduled in-office visit was protocol based [37]. In 3 (20%) of the 15 studies, only 1 scheduled in-office visit was protocol based [32-34]. Protocol-based in-office visits are described in Table 4.

Table 4. Frequencies of data transmission, review, and in-office visits of included studies.
StudyBrand of CIEDaFrequency of data transmissionFrequency of data reviewFrequency of in-office visits




TelemonitoringStandard care
Poniente [28]MedtronicContinuouslyDailyMean: 4.4 ppbMean: 7.5 pp
PREDICT RM [27]BostonNot specifiedNot specifiedPatient dependentNot specified
TARIFFc [30]St-JudeContinuouslyAfter alert and 3 monthlyAfter 12 months3 monthly
Nordland [26]BiotronikDailyAfter alertNot specifiedNot specified
EVOLVOd [11]MedtronicAfter alert and 8 monthlyAfter alert and 8 monthly8 monthly4 monthly
MORE-CAREe [29]MedtronicAfter alert and after 4, 12, and 20 monthsAfter alert and after 4, 12, and 20 months8 monthly4 monthly
Burri et al [31]BiotronikDailyAfter alert12 monthly4 monthly
Raatikainen et al [32]MedtronicAfter 3 and 6 monthsAfter 3 and 6 monthsAfter 9 months6 monthly
Al-Khatib et al [33]Medtronic3 monthly3 monthlyAfter 6 months through telephone and after 12 months3 monthly
CONNECTf [8]MedtronicAfter alert and 3 monthlyAfter alert and 3 monthlyAfter 1 month and 12 months3 monthly
ECOSTg [34]BiotronikDailyDailyAfter 1 to 3 months and thereafter 12 monthlyAfter 1 month to 3 months and thereafter 6 monthly
EuroEcoh [13]BiotronikContinuouslyDepending on the researcherAfter 6 weeks and thereafter 12 monthlyAfter 6 weeks, thereafter 12 monthly, and planned visits depending on the hospital
SAVE-HMi trial [35]Biotronik pacemakerDailyDailyNo12 monthly
SAVE-HM trial [35]Biotronik ICDjDailyDaily12 monthly6 monthly
Chew et al [36]N/AkN/AN/AN/AN/A
Dario et al [37]Biotronik, Medtronic, Boston, St-Jude, and Sorin Group pacemakerContinuouslyDailyNot unless necessary12 monthly
Dario et al [37]Biotronik, Medtronic, Boston, St-Jude, and Sorin Group ICDContinuouslyDailyAt least 16 monthly

aCIED: cardiac implantable electronic device.

bpp: per patient.

cTARIFF: Health Economics Evaluation Registry for Remote Follow-Up.

dEVOLVO: Evolution of Management Strategies of Heart Failure Patients With Implantable Defibrillators.

eMORE-CARE: Monitoring Resynchronization Devices and Cardiac Patients.

gECOST: Effectiveness and Cost of ICD Follow-Up Schedule With Telecardiology.

hEuroEco: European Health Economic Trial on Home Monitoring in ICD Patients.

iSave-HM: Socio-Economic Effects and Cost Saving Potential of Remote Patient Monitoring.

jICD: implantable-cardioverter defibrillator.

kN/A: not applicable.

Effectiveness

Effectiveness results of telemonitoring can be found in Table 2. Of the 15 studies, 9 (60%) investigated a quality-adjusted life year (QALY) or quality of life (QOL) difference [11,26-30,33,34]. A total of 53% (8/15) of studies reported an increase in QALY or QOL [11,26-28,30,33,34,39], but the QALY or QOL increase was only statistically significant in 1 (13%; P=.03) of the 8 studies [26-28,30,33,34]. In contrast, only 1 (11%) of the 9 studies investigating QOL or QALY reported a significant decrease in QOL [29]. Comparing all studies, QALY differences ranged from 0.03 to 0.27 in patients with pacemakers and ranged from −1 to 0.64 in patients with ICD or CRT-D.

Besides QALY or QOL, several studies reported other health outcomes. Chew et al [36] indicated that the risk of death was lower with telemonitoring. Al-Khatib et al [33] reported that mortality and general patient satisfaction with telemonitoring were equal to those of standard care. Crossley et al [8] reported that the time between the clinical event and the clinical decision was 17.4 days shorter in patients with an ICD or CRT-D on telemonitoring than in those on standard care (P<.001). Burri et al [31] indicated that telemonitoring patients with ICD or CRT-D led to fewer inappropriate shocks (−51%) and a reduction in battery exhaustion (−7%). Raatikainen et al [32] indicated that telemonitoring patients with an ICD reduced the average total time spent on device follow-up, with 17 minutes per patient per follow-up for physicians and 175 minutes per patient per follow-up for patients. Similarly, Dario et al [37] indicated that the time spent by physicians to treat the patient reduced by an average of 4.1 minutes per follow-up in patients with pacemakers and an average of 13.7 minutes per follow-up in patients with an ICD (SD was not reported).

Economic Impact

The results of the economic impact of telemonitoring are presented in Table 2. Of the 15 studies, 4 (27%) investigated the cost impact of telemonitoring in patients with pacemakers [26,28,35,37]. From a health care payer perspective, 1 (25%) of the 4 pacemaker studies indicated that telemonitoring increased costs with US $2183 per patient per year (not statistically significant) mainly because of increased hospitalization costs [26]. A total of 2 (50%) of the 4 pacemaker studies indicated that telemonitoring reduced costs by US $8.9 and US $1054 per patient per year mainly because of a reduction in hospitalization and staff costs, respectively [28,37]. Therefore, hospitalizations reduced costs in the study by Dario et al [37] but increased costs in the study by Lopez-Villegas et al [26]. From a patient and societal perspective, the results indicated that telemonitoring reduced costs by US $11 and US $1113 per patient per year, respectively, mainly because of lower transport costs [28,35].

Of the 15 studies, 13 (87%) investigated the cost or income impact of telemonitoring in patients with an ICD or CRT-D [8,11,13,27,29-37]. A total of 11 (85%) of the 13 ICD or CRT-D studies investigated the cost impact of telemonitoring from a health care payer perspective, all indicating that telemonitoring reduced costs for patients with an ICD or CRT-D [8,11,13,27,29-32,34,36,37]. A total of 9 (82%) of the 11 health care payer perspective studies indicated that hospitalization was the largest driver for cost reduction for patients with an ICD or CRT-D [8,11,13,27,29,30,34,36,37]. The hospitalization cost reduced by up to US $912.3 per patient per year [34]. In addition, scheduled in-office visits were reported as a driver for cost reduction in 5 (45%) of the 11 health care payer perspective studies, as up to 61% of the total cost reduction was due to a decrease in the number of scheduled in-office visits [11,29,30,32,34]. Besides cost drivers that reduced costs, there were also drivers that increased costs. In 3 (27%) of the 11 health care payer perspective studies, unscheduled visits increased the total cost impact of telemonitoring [11,13,29,30,33]. A total of 3 (20%) of the 15 studies indicated that the cost reduction for scheduled in-office visits outweighed the cost increase for unscheduled in-office visits (−US $81.4 vs US $15.6, −US $45.4 vs US $7.8, and −US $44.1 vs US $14/patient/year) [11,29,30].

The results of 4 (31%) of the 13 ICD or CRT-D studies that investigated the cost impact of telemonitoring from the patients’ perspective [11,30,32,34] indicated that patient and caregiver loss of work or activity [30], scheduled in-office visits [11], and transport [34] were the largest drivers for cost reduction. The results of 2 (15%) of the 13 ICD or CRT-D studies that investigated the income impact of telemonitoring from a provider perspective indicated that the loss of reimbursed (scheduled) in-office visits was the most important factor for income loss due to telemonitoring [13,30], reducing income by up to €72.7 (US $77.21) per patient per year [30].

ICER and Cost-Utility Ratio

Results on ICER and the cost-utility ratio are presented in Table 2. Of the 15 studies, 3 (20%) calculated the ICER from a health care payer perspective [26-28] and 1 (7%) calculated the cost-utility ratio from a health care payer perspective [11]. Of the 15 studies, 2 (13%) calculating ICER were conducted with patients with pacemakers [26,28]. Notably, of the 2 studies, 1 (50%) indicated that telemonitoring was cost-effective (ICER: US $270.09/QALY) [28], and 1 (50%) indicated that telemonitoring was not cost-effective (ICER: US $64,410/QALY) [26]. For patients with an ICD or CRT-D, of the 2 studies, 1 (50%) indicated that telemonitoring was cost-effective (ICER: US $12,069/QALY) [27] and 1 (50%) indicated that telemonitoring was dominant [11].

Critical Appraisal

The critical appraisal of the individual studies is provided in Tables 5 and 6. Of the 15 studies, 1 (7%) was classified as excellent (score of 100%) [13], 8 (53%) had a good quality score (100%<score>75%) [26,28,30,31,33,34,36,37], and 6 (40%) had a moderate quality score (75%<score>50%) [8,11,27,29,32,35]. A total of 3 (20%) of the 15 studies scored the lowest, with 59% each [8,29,32]. More than 50% (>8/15) of the studies scored low for the items cost valuation (item 9) [11,27-32,34,35,37], discounting (item 14) [8,11,29,30, 32,34,35,37], and no conflict of interest (item 18) [8,11,27, 29,30,32,34,35]. All studies scored high on the items study population (item 1), study design (item 4), time horizon (item 10), outcome identification (item 11), outcome measurement (item 12), and ethics (item 19).

Table 5. Quality assessment of the first 8 studies.

EuroEcoa [13]Nordland [26]Poniente [28]Burri et al [31]Al-Khatib et al [33]Chew et al [36]TARIFFb [30]ECOSTc [34]
Study populationd
Competing alternativesXeX
Research questionXX
Study design
Time horizon
Perspective
Cost identificationX
Cost measurement
Cost valuationXXXX
Outcome identificationX
Outcome measurement
Outcome valuationN/AfN/AN/AN/A
Incremental analysisN/AN/AN/AN/AN/AN/A
DiscountingN/AXX
Sensitivity analysisX
Conclusions
GeneralizabilityX
No conflicts of interestXX
Ethics
Values, n (%)18 (100)16 (94)15 (89)15 (88)15 (88)15 (88)15 (83)14 (82)

aEuroEco: European Health Economic Trial on Home Monitoring in ICD patients.

bTARIFF: Health Economics Evaluation Registry for Remote Follow-Up.

cECOST: Effectiveness and Cost of ICD Follow-Up Schedule With Telecardiology.

dSufficient attention was given to this aspect.

eInsufficient attention was given to this aspect.

fN/A: not applicable.

Table 6. Quality assessment of the last 7 studies.

Dario et al [37]PREDICT RM [27]SAVE-HMa [35]EVOLVOb [11]Raatikainen et al [32]MORE-CAREc [29]CONNECTd [8]
Study populatione
Competing alternativesXf
Research questionXXXXX
Study design
Time horizon
PerspectiveXX
Cost identificationXXXXX
Cost measurementXXX
Cost valuationXXXXXX
Outcome identification
Outcome measurement
Outcome valuationN/AgN/AN/AN/AN/A
Incremental analysisN/AN/AN/AN/AN/AN/A
DiscountingXXXXXX
Sensitivity analysisXXX
ConclusionsX
GeneralizabilityXX
No conflicts of interestXXXXXX
Ethics
Values, n (%)14 (82)14 (74)12 (71)12 (67)10 (59)10 (59)10 (59)

aSave-HM: Socio-Economic Effects and Cost Saving Potential of Remote Patient Monitoring.

bEVOLVO: Evolution of Management Strategies of Heart Failure Patients With Implantable Defibrillators.

cMORE-CARE: Monitoring Resynchronization Devices and Cardiac Patients.

dCONNECT: Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision.

eSufficient attention is given to this aspect.

fInsufficient attention is given to this aspect.

gN/A: not applicable.


Principal Findings and Comparison With Prior Work

The primary aim of this study was to investigate the cost-effectiveness of telemonitoring patients with an ICD or CRT-D and a pacemaker from different perspectives.

From a health care payer perspective, most studies indicated that telemonitoring was a cost-saving and effective alternative to standard care. The most important driver for cost reduction was hospitalizations, both in patients with a pacemaker and those with an ICD or CRT-D. The cost of hospitalizations was reduced by up to US $912.3 per patient per year [34]. Moreover, the reduction of scheduled in-office visits was the second most important cost-saving factor in most ICD or CRT-D studies, with up to 61% of the total cost reduction. Previous research indicated that up to 55% of the device follow-ups were routine checks with no actionable events or device programming [35,40,41]. Several researchers pointed out that most scheduled in-office visits could be replaced by telemonitoring without affecting the quality of care [7,34] and with potentially diagnosing >99.5% of arrhythmia and device problems [41]. Although scheduled in-office visits decreased, our results show that unscheduled in-office visits increased because of telemonitoring patients with an ICD or CRT-D, probably because of the possible faster detection of arrhythmia and device malfunction by telemonitoring [8]. However, in all studies analyzing both scheduled and unscheduled in-office visits, the cost reduction for scheduled in-office visits outweighed the cost increase for unscheduled in-office visits [11,29,30].

From a patient perspective, our results indicated that the reduction of professional activity, transport time, and costs due to scheduled in-office visits are the most important factors for cost reduction.

The provider perspective was investigated less frequently in the included studies, although it is very relevant. Owing to the reduction of scheduled in-office visits, providers will lose income with telemonitoring if no reimbursement exists for telemonitoring but only for in-office visits. As a result, providers will be stimulated to maintain the classic follow-up instead of telemonitoring. Of the 15 studies, 1 (7%) observed that the total cost for insurance payers does not increase in countries where telemonitoring is reimbursed [13]. As telemonitoring decreases the overall costs from a health care payer perspective, there is room for proper compensation for providers to transition from in-office care to remote care. Hence, correct compensation (which is possible while still saving on the overall health care cost) will stimulate providers to switch to telemonitoring as the desired care path for patients with a CIED.

All studies reported the effectiveness of telemonitoring. Of the 15 studies, 9 (60%) indicated a QALY or QOL difference. Furthermore, 89% (8/9) of these studies indicated an increase in QALYs or QOL for telemonitoring patients with pacemakers or ICD or CRT-D, ranging from −1 to 0.64. Some studies (3/9, 33%) indicated this QALY or QOL increase was the result of the reduced routine in-office visits [7,34]. However, the QALY or QOL increase was only statistically significant (and positive) in 1 (11%) of the 9 studies [11]. Nevertheless, patient questionnaires have demonstrated a high acceptance of telemonitoring among patients with pacemakers and those with ICDs [39]. Moreover, telemonitoring is reported to lead to an increased sense of security [39]. Furthermore, the results indicated that telemonitoring leads to fewer inappropriate shocks, an important determinant of QALY, in patients with an ICD or CRT-D [31].

The cost-effectiveness analyses may be sensitive to the heterogeneity among the organization of telemonitoring in different hospitals. This may include different devices, the number of transmissions, the configuration of alerts, and hospital visit scheduling [26]. It seems reasonable to expect that the efficiency of telemonitoring not only depends on the technology but also on the organization of the service. If hospitals see telemonitoring as an additional service, on top of standard care, less cost-savings may be seen than if hospitals see telemonitoring as a substitute for standard care. A radical organizational change could lead to larger cost-savings, as suggested by an observational study by Facchin et al [42]. Moreover, such radical change may include a strategy involving other physicians, such as general practitioners, and referring cardiologists, that is, an integrated health care delivery [37].

Furthermore, the comparison between studies is challenged by differences in study design. The Poniente study by Bautista-Mesa et al [28] followed up patients with pacemakers for 12 months and indicated a QALY increase of 0.09 for telemonitoring. However, after 5 years of follow-up, the results indicated a QALY decrease of 0.20 for telemonitoring. Bautista-Mesa et al [28] indicated that some of the telemonitoring benefits (eg, reduction of in-office visits) may not be appreciated in the long term. Therefore, the evolution of utilities may be different depending on the follow-up time. In addition, the results indicated that hospitalizations reduced costs in the study by Dario et al [37] but increased costs in the study by Lopez-Villegas et al [26]. This discrepancy might be explained because significantly fewer patients were included in the study by Lopez-Villegas et al (50 vs 2101 patients). None of the 25 patients in the conventional follow-up group were hospitalized, whereas 12% (3/25) of the patients were hospitalized in the remotely monitored group (all for pacemaker problems) [26]. Furthermore, the included studies relied disproportionally on male participants, except for the Nordland study [26]. This may be explained by the significant sex disparity in ICD implantation rates, pointed out by Ingelaere et al [43]. Ingelaere et al [43] could not completely explain these differences by prevalence differences of cardiomyopathies and imply a possible undertreatment of women. Another study [44] observed an undertreatment of women with coronary heart disease, as they are less likely to undergo coronary angiography. Therefore, men may undergo more expensive treatments than women. This can explain why the included cost-effectiveness studies may present an overly positive result. In addition, time differences may impact the quality and cost-effectiveness of telemonitoring, as telemonitoring may evolve over time. However, our results did not provide meaningful insights in this respect.

The cost-effectiveness analyses may be sensitive to the heterogeneity among health care systems. From a provider perspective, our results indicated that telemonitoring generates lesser profit than standard care in the absence of reimbursement. Therefore, the lack of reimbursement is generally perceived as a major implementation barrier to telemonitoring, affecting 80% of the centers [45]. Consequently, providers tend to continue with standard care instead of telemonitoring. However, from a health care payer perspective, our results indicated that telemonitoring was still cost-saving even with reimbursement [13,34]. To stimulate providers to use telemonitoring, provider compensation should be provided based on overall health care cost-savings, making telemonitoring possible if it is preferred as the way to deliver CIED follow-up care.

Limitations

Because of the large discrepancies between health care systems’ organization, costs, access, delivery, quality, and reimbursement of cardiac care, any generalization may be perceived as inaccurate [37,46]. For instance, the included studies were mainly performed in Western countries. The results may not be generalizable to non-Western countries. Therefore, the cost-effectiveness results are contingent on the context in which they were analyzed [46]. Another limitation of this research is that 40% (6/15) of the included studies are not randomized controlled trials. These studies may have unobserved confounding factors that cannot be controlled for. Finally, cost analyses were excluded in this study because of our research objective. However, future cost analyses could draw a lot of information from analyzing these excluded studies.

Conclusions

Telemonitoring patients with CIED may be a cost-effective alternative to standard follow-up. Moreover, telemonitoring may lead to a cost reduction from a health care and patient perspective, mainly by the reduction of hospitalizations and scheduled in-office visits. Owing to the reduction in scheduled in-office visits, providers’ income tends to decrease when implementing telemonitoring without proper reimbursement. Introducing appropriate reimbursement could make telemonitoring sustainable for providers, while still being cost-effective from a health care payer perspective.

Acknowledgments

The authors would like to thank Dr Ingrid Kremer of Maastricht University for her help in the manuscript review. This work was supported by the Fund for Scientific Research Flanders (Fonds Wetenschappelijk Onderzoek Vlaanderen, grant 1SC9322N, 2021). RW is supported as a postdoctoral clinical researcher by the Fund for Scientific Research Flanders (Fonds Wetenschappelijk Onderzoek Vlaanderen).

Data Availability

All data generated or analyzed during this study are included in this published article and its multimedia appendices.

Authors' Contributions

SR and LA were responsible for the conceptualization of the manuscript. SR, LA, RW, and HH acquired the financial support necessary for this paper and developed the methodology. SR analyzed and investigated the data. AP, LA, RW, and HH validated the results. SR was responsible for the first and final drafts. LA, RW, and HH were involved in editing the drafts. All authors approved the final manuscript.

Conflicts of Interest

RW reports research funding from Abbott, Biotronik, Boston Scientific, and Medtronic and speakers and consultancy fees from Medtronic, Boston Scientific, Biotronik, and Abbott. None of these payments were personal; all were handled through the University of Leuven. HH received personal lecture and consultancy fees from Abbott, Biotronik, Daiichi-Sankyo, Pfizer-BMS, Medscape, and Springer Healthcare Limited. He received unconditional research grants through the University of Antwerp and the University of Hasselt from Abbott, Bayer, Biotronik, Biosense Webster, Boston Scientific, Boehringer Ingelheim, Daicchi-Sankyo, Fibricheck or Qompium, Medtronic, and Pfizer-BMS, all outside the scope of this work. All other authors declare no other conflicts of interest.

Multimedia Appendix 1

PRISMA-ScR (Preferred Reporting Items for Systematic Reviews and Meta-Analyses extension for Scoping Reviews) checklist.

DOCX File , 108 KB

Multimedia Appendix 2

Search strategy.

DOCX File , 14 KB

Multimedia Appendix 3

Interpretation Consensus Health Economic Criteria list.

DOCX File , 18 KB

  1. Blomström-Lundqvist C, Traykov V, Erba PA, Burri H, Nielsen JC, Bongiorni MG, et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections-endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID) and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Europace. Apr 01, 2020;22(4):515-549. [FREE Full text] [CrossRef] [Medline]
  2. Wilkoff BL, Auricchio A, Brugada J, Cowie M, Ellenbogen KA, Gillis AM, et al. HRS/EHRA expert consensus on the monitoring of cardiovascular implantable electronic devices (CIEDs): description of techniques, indications, personnel, frequency and ethical considerations. Heart Rhythm. Jun 2008;5(6):907-925. [CrossRef] [Medline]
  3. de la Torre Díez I, Garcia-Zapirain B, Méndez-Zorrilla A, López-Coronado M. Monitoring and follow-up of chronic heart failure: a literature review of eHealth applications and systems. J Med Syst. Jul 2016;40(7):179. [CrossRef] [Medline]
  4. Guédon-Moreau L, Lacroix D, Sadoul N, Clémenty J, Kouakam C, Hermida J, et al. A randomized study of remote follow-up of implantable cardioverter defibrillators: safety and efficacy report of the ECOST trial. Eur Heart J. Feb 2013;34(8):605-614. [FREE Full text] [CrossRef] [Medline]
  5. Ricci RP, Morichelli L, Santini M. Remote control of implanted devices through Home Monitoring technology improves detection and clinical management of atrial fibrillation. Europace. Jan 2009;11(1):54-61. [FREE Full text] [CrossRef] [Medline]
  6. García-Fernández FJ, Osca Asensi J, Romero R, Fernández Lozano I, Larrazabal JM, Martínez Ferrer J, et al. Safety and efficiency of a common and simplified protocol for pacemaker and defibrillator surveillance based on remote monitoring only: a long-term randomized trial (RM-ALONE). Eur Heart J. Jun 14, 2019;40(23):1837-1846. [FREE Full text] [CrossRef] [Medline]
  7. Varma N, Epstein AE, Irimpen A, Schweikert R, Love C. Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator follow-up: the Lumos-T Safely Reduces Routine Office Device Follow-up (TRUST) trial. Circulation. Jul 27, 2010;122(4):325-332. [FREE Full text] [CrossRef] [Medline]
  8. Crossley GH, Boyle A, Vitense H, Chang Y, Mead RH. The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) trial: the value of wireless remote monitoring with automatic clinician alerts. J Am Coll Cardiol. Mar 8, 2011;57(10):1181-1189. [FREE Full text] [CrossRef] [Medline]
  9. Catalan-Matamoros D, Lopez-Villegas A, Tore-Lappegard K, Lopez-Liria R. Patients' experiences of remote communication after pacemaker implant: The NORDLAND study. PLoS One. 2019;14(6):e0218521. [FREE Full text] [CrossRef] [Medline]
  10. Masella C, Zanaboni P, Di Stasi F, Gilardi S, Ponzi P, Valsecchi S. Assessment of a remote monitoring system for implantable cardioverter defibrillators. J Telemed Telecare. 2008;14(6):290-294. [CrossRef] [Medline]
  11. Zanaboni P, Landolina M, Marzegalli M, Lunati M, Perego GB, Guenzati G, et al. Cost-utility analysis of the EVOLVO study on remote monitoring for heart failure patients with implantable defibrillators: randomized controlled trial. J Med Internet Res. 2013;15(5):e106. [FREE Full text] [CrossRef] [Medline]
  12. Annemans L. Health Economics for Non-economists: An Introduction to the Concepts, Methods and Pitfalls of Health Economic Evaluations. Kalmthout, Belgium. Pelckmans Pro; 2018.
  13. Heidbuchel H, Hindricks G, Broadhurst P, Van Erven L, Fernandez-Lozano I, Rivero-Ayerza M, et al. EuroEco (European Health Economic Trial on Home Monitoring in ICD Patients): a provider perspective in five European countries on costs and net financial impact of follow-up with or without remote monitoring. Eur Heart J. Jan 14, 2015;36(3):158-169. [FREE Full text] [CrossRef] [Medline]
  14. Klersy C, Boriani G, De Silvestri A, Mairesse GH, Braunschweig F, Scotti V, et al. Effect of telemonitoring of cardiac implantable electronic devices on healthcare utilization: a meta-analysis of randomized controlled trials in patients with heart failure. Eur J Heart Fail. Feb 2016;18(2):195-204. [FREE Full text] [CrossRef] [Medline]
  15. Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. Mar 29, 2021;372:n71. [FREE Full text] [CrossRef] [Medline]
  16. Tricco AC, Lillie E, Zarin W, O'Brien KK, Colquhoun H, Levac D, et al. PRISMA extension for scoping reviews (PRISMA-ScR): checklist and explanation. Ann Intern Med. Oct 02, 2018;169(7):467-473. [FREE Full text] [CrossRef] [Medline]
  17. Wijnen B, Van Mastrigt G, Redekop WK, Majoie H, De Kinderen R, Evers S. How to prepare a systematic review of economic evaluations for informing evidence-based healthcare decisions: data extraction, risk of bias, and transferability (part 3/3). Expert Rev Pharmacoecon Outcomes Res. Dec 2016;16(6):723-732. [CrossRef] [Medline]
  18. Search strategies: the centre for reviews and dissemination. University of York. 2018. URL: http://www.crd.york.ac.uk/crdweb/searchstrategies.asp [accessed 2024-07-01]
  19. Auener SL, Remers TE, van Dulmen SA, Westert GP, Kool RB, Jeurissen PP. The effect of noninvasive telemonitoring for chronic heart failure on health care utilization: systematic review. J Med Internet Res. Sep 29, 2021;23(9):e26744. [FREE Full text] [CrossRef] [Medline]
  20. Lopez-Villegas A, Leal-Costa C, Perez-Heredia M, Villegas-Tripiana I, Catalán-Matamoros D. Knowledge update on the economic evaluation of pacemaker telemonitoring systems. Int J Environ Res Public Health. Nov 18, 2021;18(22):12120. [FREE Full text] [CrossRef] [Medline]
  21. Vehmeijer JT, Brouwer TF, Limpens J, Knops RE, Bouma BJ, Mulder BJ, et al. Implantable cardioverter-defibrillators in adults with congenital heart disease: a systematic review and meta-analysis. Eur Heart J. May 07, 2016;37(18):1439-1448. [FREE Full text] [CrossRef] [Medline]
  22. Scott AM, Forbes C, Clark J, Carter M, Glasziou P, Munn Z. Systematic review automation tools improve efficiency but lack of knowledge impedes their adoption: a survey. J Clin Epidemiol. Oct 2021;138:80-94. [CrossRef] [Medline]
  23. Evers S, Goossens M, de VH, van TM, Ament A. Criteria list for assessment of methodological quality of economic evaluations: consensus on Health Economic Criteria. Int J Technol Assess Health Care. 2005;21(2):240-245. [Medline]
  24. Jiang X, Ming W, You JH. The cost-effectiveness of digital health interventions on the management of cardiovascular diseases: systematic review. J Med Internet Res. Jun 17, 2019;21(6):e13166. [FREE Full text] [CrossRef] [Medline]
  25. Consumptieprijsindex. Statbel. URL: https://statbel.fgov.be/nl/themas/consumptieprijsindex/consumptieprijsindex#figures [accessed 2022-07-08]
  26. Lopez-Villegas A, Catalan-Matamoros D, Peiro S, Lappegard KT, Lopez-Liria R. Cost-utility analysis of telemonitoring versus conventional hospital-based follow-up of patients with pacemakers. The NORDLAND randomized clinical trial. PLoS One. 2020;15(1):e0226188. [FREE Full text] [CrossRef] [Medline]
  27. Hummel JP, Leipold RJ, Amorosi SL, Bao H, Deger KA, Jones PW, et al. Outcomes and costs of remote patient monitoring among patients with implanted cardiac defibrillators: an economic model based on the PREDICT RM database. J Cardiovasc Electrophysiol. Jul 2019;30(7):1066-1077. [FREE Full text] [CrossRef] [Medline]
  28. Bautista-Mesa RJ, Lopez-Villegas A, Peiro S, Catalan-Matamoros D, Robles-Musso E, Lopez-Liria R, et al. Long-term cost-utility analysis of remote monitoring of older patients with pacemakers: the PONIENTE study. BMC Geriatr. Nov 16, 2020;20(1):474. [FREE Full text] [CrossRef] [Medline]
  29. Boriani G, Da Costa A, Quesada A, Ricci RP, Favale S, Boscolo G, et al. Effects of remote monitoring on clinical outcomes and use of healthcare resources in heart failure patients with biventricular defibrillators: results of the MORE-CARE multicentre randomized controlled trial. Eur J Heart Fail. Mar 2017;19(3):416-425. [FREE Full text] [CrossRef] [Medline]
  30. Ricci RP, Vicentini A, D'Onofrio A, Sagone A, Rovaris G, Padeletti L, et al. Economic analysis of remote monitoring of cardiac implantable electronic devices: results of the health economics evaluation registry for remote follow-up (TARIFF) study. Heart Rhythm. Jan 2017;14(1):50-57. [FREE Full text] [CrossRef] [Medline]
  31. Burri H, Sticherling C, Wright D, Makino K, Smala A, Tilden D. Cost-consequence analysis of daily continuous remote monitoring of implantable cardiac defibrillator and resynchronization devices in the UK. Europace. Nov 2013;15(11):1601-1608. [FREE Full text] [CrossRef] [Medline]
  32. Raatikainen MJ, Uusimaa P, van Ginneken MM, Janssen JP, Linnaluoto M. Remote monitoring of implantable cardioverter defibrillator patients: a safe, time-saving, and cost-effective means for follow-up. Europace. Oct 2008;10(10):1145-1151. [FREE Full text] [CrossRef] [Medline]
  33. Al-Khatib SM, Piccini JP, Knight D, Stewart M, Clapp-Channing N, Sanders GD. Remote monitoring of implantable cardioverter defibrillators versus quarterly device interrogations in clinic: results from a randomized pilot clinical trial. J Cardiovasc Electrophysiol. May 2010;21(5):545-550. [CrossRef] [Medline]
  34. Guédon-Moreau L, Lacroix D, Sadoul N, Clémenty J, Kouakam C, Hermida J, et al. Costs of remote monitoring vs. ambulatory follow-ups of implanted cardioverter defibrillators in the randomized ECOST study. Europace. Aug 2014;16(8):1181-1188. [FREE Full text] [CrossRef] [Medline]
  35. Perl S, Stiegler P, Rotman B, Prenner G, Lercher P, Anelli-Monti M, et al. Socio-economic effects and cost saving potential of remote patient monitoring (SAVE-HM trial). Int J Cardiol. Nov 30, 2013;169(6):402-407. [CrossRef] [Medline]
  36. Chew DS, Zarrabi M, You I, Morton J, Low A, Reyes L, et al. Clinical and economic outcomes associated with remote monitoring for cardiac implantable electronic devices: a population-based analysis. Can J Cardiol. Jun 2022;38(6):736-744. [CrossRef] [Medline]
  37. Dario C, Delise P, Gubian L, Saccavini C, Brandolino G, Mancin S. Large controlled observational study on remote monitoring of pacemakers and implantable cardiac defibrillators: a clinical, economic, and organizational evaluation. Interact J Med Res. Jan 13, 2016;5(1):e4. [FREE Full text] [CrossRef] [Medline]
  38. Crossley G, Boyle A, Vitense H, Sherfesee L, Mead RH. Trial design of the clinical evaluation of remote notification to reduce time to clinical decision: the clinical evaluation Of remote NotificatioN to rEduCe Time to clinical decision (CONNECT) study. Am Heart J. Nov 2008;156(5):840-846. [CrossRef] [Medline]
  39. Ricci RP, Morichelli L, Quarta L, Sassi A, Porfili A, Laudadio MT, et al. Long-term patient acceptance of and satisfaction with implanted device remote monitoring. Europace. May 2010;12(5):674-679. [CrossRef] [Medline]
  40. Boriani G, Auricchio A, Klersy C, Kirchhof P, Brugada J, Morgan J, et al. Healthcare personnel resource burden related to in-clinic follow-up of cardiovascular implantable electronic devices: a European Heart Rhythm Association and Eucomed joint survey. Europace. Aug 2011;13(8):1166-1173. [CrossRef] [Medline]
  41. Heidbüchel H, Lioen P, Foulon S, Huybrechts W, Ector J, Willems R, et al. Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator. Europace. Mar 2008;10(3):351-357. [FREE Full text] [CrossRef] [Medline]
  42. Facchin D, Baccillieri MS, Gasparini G, Zoppo F, Allocca G, Brieda M, et al. Findings of an observational investigation of pure remote follow-up of pacemaker patients: is the in-clinic device check still needed? Int J Cardiol. Oct 01, 2016;220:781-786. [CrossRef] [Medline]
  43. Ingelaere S, Hoffmann R, Guler I, Vijgen J, Mairesse GH, Blankoff I, et al. Inequality between women and men in ICD implantation. Int J Cardiol Heart Vasc. Aug 2022;41:101075. [FREE Full text] [CrossRef] [Medline]
  44. Steingart RM, Packer M, Hamm P, Coglianese ME, Gersh B, Geltman EM, et al. Sex differences in the management of coronary artery disease. Survival and Ventricular Enlargement Investigators. N Engl J Med. Jul 25, 1991;325(4):226-230. [CrossRef] [Medline]
  45. Mairesse GH, Braunschweig F, Klersy K, Cowie MR, Leyva F. Implementation and reimbursement of remote monitoring for cardiac implantable electronic devices in Europe: a survey from the health economics committee of the European Heart Rhythm Association. Europace. May 2015;17(5):814-818. [CrossRef] [Medline]
  46. Lucà F, Cipolletta L, Di Fusco SA, Iorio A, Pozzi A, Rao CM, et al. Remote monitoring: doomed to let down or an attractive promise? IJC Heart Vasculature. Sep 2019;24:100380. [CrossRef]


CHEC: Consensus Health Economic Criteria
CIED: cardiac implantable electronic device
CRT-D: cardiac resynchronization therapy defibrillator
ICD: implantable cardioverter-defibrillator
ICER: incremental cost-effectiveness ratio
PRISMA: Preferred Reporting Items for Systematic Reviews and Meta-Analyses
PRISMA-ScR: Preferred Reporting Items for Systematic Reviews and Meta-Analyses extension for Scoping Reviews
PROSPERO: International Prospective Register of Systematic Reviews
QALY: quality-adjusted life year
QOL: quality of life


Edited by T Leung; submitted 28.03.23; peer-reviewed by P Jeurissen, B Dechert; comments to author 07.09.23; revised version received 13.09.23; accepted 13.02.24; published 19.04.24.

Copyright

©Sarah Raes, Andrea Prezzi, Rik Willems, Hein Heidbuchel, Lieven Annemans. Originally published in the Journal of Medical Internet Research (https://www.jmir.org), 19.04.2024.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research, is properly cited. The complete bibliographic information, a link to the original publication on https://www.jmir.org/, as well as this copyright and license information must be included.