Published on in Vol 19, No 5 (2017): May

Adherence to AHA Guidelines When Adapted for Augmented Reality Glasses for Assisted Pediatric Cardiopulmonary Resuscitation: A Randomized Controlled Trial

Adherence to AHA Guidelines When Adapted for Augmented Reality Glasses for Assisted Pediatric Cardiopulmonary Resuscitation: A Randomized Controlled Trial

Adherence to AHA Guidelines When Adapted for Augmented Reality Glasses for Assisted Pediatric Cardiopulmonary Resuscitation: A Randomized Controlled Trial

Original Paper

1Geneva Children’s Hospital, Department of Pediatric Emergency Medicine, University Hospitals of Geneva, Geneva, Switzerland

2Division of Medical Information Sciences, Department of Radiology and Medical Informatics, University Hospitals of Geneva, Geneva, Switzerland

3Geneva Medical Center, University Hospitals of Geneva, Geneva, Switzerland

*these authors contributed equally

Corresponding Author:

Johan N Siebert, MD

Geneva Children’s Hospital

Department of Pediatric Emergency Medicine

University Hospitals of Geneva

Avenue de la Roseraie, 47

Geneva, 1205

Switzerland

Phone: 41 795534072

Fax:41 223725423

Email: Johan.Siebert@hcuge.ch


Background: The American Heart Association (AHA) guidelines for cardiopulmonary resuscitation (CPR) are nowadays recognized as the world’s most authoritative resuscitation guidelines. Adherence to these guidelines optimizes the management of critically ill patients and increases their chances of survival after cardiac arrest. Despite their availability, suboptimal quality of CPR is still common. Currently, the median hospital survival rate after pediatric in-hospital cardiac arrest is 36%, whereas it falls below 10% for out-of-hospital cardiac arrest. Among emerging information technologies and devices able to support caregivers during resuscitation and increase adherence to AHA guidelines, augmented reality (AR) glasses have not yet been assessed. In order to assess their potential, we adapted AHA Pediatric Advanced Life Support (PALS) guidelines for AR glasses.

Objective: The study aimed to determine whether adapting AHA guidelines for AR glasses increased adherence by reducing deviation and time to initiation of critical life-saving maneuvers during pediatric CPR when compared with the use of PALS pocket reference cards.

Methods: We conducted a randomized controlled trial with two parallel groups of voluntary pediatric residents, comparing AR glasses to PALS pocket reference cards during a simulation-based pediatric cardiac arrest scenario—pulseless ventricular tachycardia (pVT). The primary outcome was the elapsed time in seconds in each allocation group, from onset of pVT to the first defibrillation attempt. Secondary outcomes were time elapsed to (1) initiation of chest compression, (2) subsequent defibrillation attempts, and (3) administration of drugs, as well as the time intervals between defibrillation attempts and drug doses, shock doses, and number of shocks. All these outcomes were assessed for deviation from AHA guidelines.

Results: Twenty residents were randomized into 2 groups. Time to first defibrillation attempt (mean: 146 s) and adherence to AHA guidelines in terms of time to other critical resuscitation endpoints and drug dose delivery were not improved using AR glasses. However, errors and deviations were significantly reduced in terms of defibrillation doses when compared with the use of the PALS pocket reference cards. In a total of 40 defibrillation attempts, residents not wearing AR glasses used wrong doses in 65% (26/40) of cases, including 21 shock overdoses >100 J, for a cumulative defibrillation dose of 18.7 Joules per kg. These errors were reduced by 53% (21/40, P<.001) and cumulative defibrillation dose by 37% (5.14/14, P=.001) with AR glasses.

Conclusions: AR glasses did not decrease time to first defibrillation attempt and other critical resuscitation endpoints when compared with PALS pocket cards. However, they improved adherence and performance among residents in terms of administering the defibrillation doses set by AHA.

J Med Internet Res 2017;19(5):e183

doi:10.2196/jmir.7379

Keywords



Clinical practice guidelines aim to improve quality of care, reduce variation of practice, and provide evidence-based health care [1]. The American Heart Association (AHA) guidelines for cardiopulmonary resuscitation (CPR) are nowadays recognized as the world’s most authoritative resuscitation guidelines [2,3]. They are evidence-based, synthesized by experts, and include a large number of algorithms intended to provide step-by-step processes to various life-threatening emergency situations in a systematic fashion. These algorithms are also summarized on pocket reference cards in order to be used as quick reference tools that emergency physicians may have access to during resuscitations. However, despite their availability, suboptimal quality of resuscitation is still common for both adult and pediatric patients [4]. Immediate (level 1) triage represents 175,000 patient visits every year in US pediatric emergency departments (PED) [5]. Among them, 5800 to 10,000 cases are due to in-hospital cardiac arrest (IHCA) [6,7], and 6700 to 15,000 cases to out-of-hospital cardiac arrest (OHCA) [8-10], including 6000 related to non-traumatic causes [11]. Quality CPR with adherence to AHA resuscitation guidelines optimizes the management of critically ill patients and increases their chances of survival [12,13], whereas deviation is associated with decreased likelihood of survival from cardiac arrest (CA) [14]. Currently, the median hospital survival rate from pediatric IHCA is 36% [4], whereas it is below 10% for OHCA [15,16].

As a result, the scientific community has proposed new resuscitation strategies relying on information technologies and devices aiming at improving and ensuring adherence to AHA guidelines [17-20]. Among possible emerging information technologies that could support caregivers, augmented reality (AR) glasses have recently gained a great deal of interest within the scientific community. AR glasses are wearable and connected devices that display interactive images to the visual field of users by overlaying visual information without significantly disturbing the ordinary vision. They feature some functionalities similar to those offered by smartphones and tablets by running self-contained mobile apps. Despite recent communications and studies related to the use of these glasses in various medical fields [21-26], their contribution to resuscitation in emergency medicine has not yet been investigated. Their potential to wirelessly display and interact in real time conditions with data generated by the actions performed in a specific environment could be of great interest for assisting caregivers in resuscitation rooms, while freeing users’ hands and allowing them to “see the scene through the screen”. Augmented reality might bring useful information to caregivers’ attention without disturbing the care process, by allowing constant access to the resuscitation scene. In a previous article, we have described the adaptation of AHA Pediatric Advanced Life Support (PALS) algorithms for AR glasses [27]. This study aimed at investigating whether this adaptation for AR glasses would increase adherence to AHA guidelines by reducing deviation and time to initiation of critical life-saving maneuvers during pediatric CPR when compared with the use of PALS pocket reference cards, in a simulated model.


Study Design

We conducted a prospective, randomized controlled trial with 2 parallel groups of voluntary pediatric residents comparing time to first defibrillation attempt while using AR glasses (Google Glass, allocation group A) or AHA PALS conventional pocket reference cards (allocation group B) during a standardized simulation-based pediatric CA scenario (Multimedia Appendix 1). No changes were made to the AR glasses or the intervention during the study.

Selection of Participants

Any physician from the whole pediatric department actively training for a pediatric specialty (residents) was eligible. Residents with eye disorders were not included in the study. Shift-working residents were randomly recruited on the day of the study from a random alphabetical list. Written informed consent was obtained from all the participants before their voluntary involvement. Study participants were not involved in the study design, choice of outcome measures, or the execution of the study. No participants were asked for advice on interpretation or the writing of results. The results of the study were offered to the study participants after the completion of the study.

Setting and Resuscitation Scenario

The study was conducted in a pediatric emergency department (PED) of a tertiary hospital with approximately 28,000 visits per year. We created a standardized simulation scenario on a high-fidelity manikin (Laerdal SimJunior). The resuscitation team was composed of the resident participating in the study and 3 nurses to assist with resuscitation through drugs preparation, chest compressions, and bag-valve-mask ventilation, according to the resident’s instructions. A certified technician (KH), who was not a member of the resuscitation team, operated the simulator. Except for participating residents, members of the resuscitation team remained unchanged across all scenarios and were the investigators in the study.

Figure 1. American Heart Association’s pediatric cardiac arrest algorithm—2015 update.
View this figure
Figure 2. Screenshot of a resuscitation step from the American Heart Association’s Pediatric Advanced Life Support pulseless ventricular tachycardia algorithm as adapted in augmented reality glasses.
View this figure

On the day of participation, residents were asked about their demographics. After random allocation, each participant allocated to the AR glasses group received a standardized 15-minute qualifying training session to familiarize them uniformly with the AR glasses. Then, all the participants were asked to perform a 15-minute highly realistic CPR scenario (Multimedia Appendices 2 and 3). Unlike adults, CA in children without prior cardiac disease is mainly due to asystole (40%) and pulseless electrical activity (24%) [28]. Ventricular fibrillation and pulseless ventricular tachycardia (pVT), namely shockable rhythms, are identified in 27% of pediatric IHCA [29]. We decided to study the pVT algorithm because, in our opinion, it offers a greater opportunity to assess the multiple-steps resuscitative skills set by AHA. The scenario was therefore standardized to strictly follow the 2015 AHA pediatric pVT algorithm (Figure 1) [30] and provided on the same manikin. It was conducted in situ in the pediatric resuscitation room of the PED to increase realism. No interactions occurred between participants and investigators. When entering the room, a short clinical statement to recognize the life-threatening condition of the patient, including his weight and age, was given to the resident. The resident was then asked to start the timed scenario and had to recognize by himself or herself the previously settled cardiac rhythm (pVT) with the AR glasses (allocation group A) or the PALS conventional pocket reference card (group B). All participants in allocation group B were required to have their PALS pocket card in their hands throughout the entire scenario. Whether they referred to it or not was left to them, as in real life. The scenario ran invariably until the manikin was defibrillated at the 4th shock and showed a subsequent return of spontaneous circulation (ROSC). In order to be consistent with the 2015 AHA pediatric CA algorithm [30] and to standardize our scenario, defibrillation doses of 2 Joules per kg for the first attempt, and 4 Joules per kg for subsequent 2nd, 3rd, and 4th attempts were expected (Figure 1).

Intervention

As previously described [27], the numerous steps of AHA PALS algorithms were wisely split into “cards.” Each card transposed to the AR glasses paralleled the informational content of a resuscitation step from the original algorithm. However, the informational content was “augmented,” thanks to the interactivity and display capability of the device. The AR algorithm thus obtained was set up in a manner similar to the PALS pocket references regarding the progression and sequence of actions along the original algorithm’s sequences. For instance, the complete pulseless ventricular tachycardia (pVT) algorithm was adapted on 42 cards designed to be as concise as possible without hindering proper progression along the algorithm. The cards were tailored to the small size of the Google glass screen, following a user-centered and ergonomic-driven approach. Each card was structured on 4 zones: (1) a color-coded title allowing direct identification of each step in progress, (2) an image on the left helping with decision-making (such as distinctive illustration of cardiac rhythms), (3) a menu choice on the right helping to progress in the resuscitation steps, and (4) a footer to preview the next step (Figure 2). Interaction was also defined with end users. Tactile commands of the glasses were favored over voice commands due to the inability of AR glasses to distinguish between vocal orders in the noisy environment of a resuscitation room. Swiping up or down allowed navigating inside the card. Selection was done by a click and actions cancelled by swiping back. Each cycle of chest compression–ventilation was timed, thanks to a countdown clock displayed on the screen. Treatment and defibrillation doses (Philips HeartStart MRx Biphasic Defibrillator) were automatically calculated on patients’ weight or age.

Outcome Measures

The elapsed time in seconds in each allocation group from onset of pulseless shockable rhythm to the first defibrillation attempt was selected as the primary outcome, as it is the most important determinant of survival after CA [31]. Secondary outcomes were time elapsed to initiation of chest compression, time to subsequent defibrillation attempts, time to administration of epinephrine and amiodarone, and time interval (in seconds) between defibrillation attempts. AHA recommends 5 cycles of chest compression (about 2 minutes) between defibrillation attempts. The amount of time spent by participants to perform chest compressions by cycles of chest compression was defined as the hands-on time. It was measured in seconds with a chronometer. Drug doses, shock doses, and number of shocks were also assessed. All these outcomes were assessed for deviation from AHA guidelines. At the end of the scenario, a questionnaire using a 10-point Likert scale was submitted to the participant to measure the overall stress perceived during the scenario.

Methods of Measurement and Data Collection

All the actions (ie, the primary and secondary outcomes) performed by the resident during the scenario were automatically recorded and stored by the responsive simulator detectors and by several video cameras. The videos were embedded in a dedicated simulation software, allowing accurate assessment of timing and sequencing of actions. To avoid assessment bias, 2 evaluators then independently reviewed these video recordings. In case of disagreement, a third independent evaluator helped reach a consensus. Data were manually retrieved and entered into a Microsoft Excel spreadsheet (version 2011). Unaccomplished actions were left blank and not assigned any corresponding time. Only residents were assessed and their privacy preserved. Only the investigators of the study had access to the data. The statistical software GraphPad Prism version 6.0h (GraphPad Software, Inc) was used for all data analyses.

Sample Size

The primary objective of the study was to detect a difference in time to the first defibrillation attempt. The sample size was calculated to detect a 30-second decrease in time to first defibrillation between 2 independent groups with a power of 80% and a 2-sided risk alpha of .05. A previous study has shown a mean time to first defibrillation of 92 seconds [32] with a standard deviation (SD) of 23 seconds. Assuming a similar SD in each group in our study, 10 patients per group were required.

Randomization and Blinding

We randomly assigned residents in a 1:1 ratio with a Web-based software [33]. Blinding to the purpose of the study during recruitment was maintained to minimize preparation bias. Participants were unblinded after randomization. Allocation concealment was ensured with the same Web-based software [33] and was not released until the residents started the scenario.

Statistical Analysis
Primary Outcome

We first evaluated the time elapsed between onset of pVT and first defibrillation attempt. The Shapiro-Wilks test was used for normality analysis of the parameters. Means and standard deviations (SDs) with 95% CI were reported. Non-normally distributed variables were analyzed using the Mann-Whitney test. Frequencies were reported as percentages. T tests were used to compare independent groups. No paired data were compared. Kaplan–Meier curves for time elapsed between onset of pVT and 1) initiation of chest compression, and 2) first defibrillation, were estimated and compared using the log-rank (Mantel-Cox) test for bivariate survival analysis.

Secondary Outcomes

We evaluated the time elapsed between onset of pVT to subsequent defibrillation attempts and drug delivery. As most of the continuous variables were also normally distributed, means and SDs with 95% CI were reported. Non-normally distributed variables were analyzed using a Mann-Whitney test. Frequencies were reported as percentages. T tests were used to compare independent groups. No paired data were compared. Errors in cycles of chest compression-ventilation were measured as the deviation in percent from the experimental time spent in seconds compared with the 2-minute duration recommended by AHA guidelines. Wrong defibrillation or drug doses were measured as the deviation in percent from the amount of energy delivered in Joules or drug doses in milliliters compared with AHA recommendations. Wrong defibrillation mode was also measured. Absolute deviations were also analyzed. The mean (SD) difference in deviation obtained with each method was reported with 95% CI. A t test for unpaired data was used to compare interventions. Mean differences were reported by randomized group. We also determined if prior certification as a PALS provider before the study, prior resuscitations as a provider in real-life, or post-graduation years (PGY) had a significant impact on the above outcomes. Mean and SD were determined for stress scores of individuals in the questionnaire and reported with descriptive statistics. A P value less than .05 was considered significant.

Ethics and Informed Consent

The study was approved by the institutional ethics committee. According to the ICMJE, a registration number was not required for our trial, as the purpose of this study was to examine the effect of the intervention on the providers. Written informed consent was obtained from all participants before their voluntary involvement. The study was conducted in accordance with the principles of the Declaration of Helsinki, the standards of Good Clinical Practice, and Swiss regulatory requirements.

Figure 3. CONSORT flowchart of augmented reality glasses trial.
View this figure

Study Participants

In March 2016, 20 pediatric residents participated and completed the study with no dropout (Figure 3). The demographic results are summarized in Table 1.

Time to Resuscitation Critical Endpoints

The first defibrillation was delivered within 180 seconds by 70% of the residents in both groups (Figure 4). Mean times to resuscitation critical endpoints are summarized in Table 2. None of them were significantly different between allocation groups A and B. All participants (100%) correctly recognized the pVT rhythm on the monitor. In both allocation groups, 90% of the residents initiated chest compressions within 60 seconds from the onset of pVT, and half of them before 20 seconds. There were no statistically significant differences in hands-on time spent by cycles of chest compression between both groups. Due to a lack of some defibrillation attempts, 1 interval in allocation group A and 4 in group B were not measurable.

At the time of the study, 6 participants (60%) in allocation group A and 7 participants (70%) in allocation group B were PALS-certified providers. Eleven participants out of 20 (55%) were residents with more than 1 year of PGY. With regard to all outcomes measured, we observed in both allocation groups that PALS-certified residents or those with PGY>1 tended to defibrillate and deliver drugs more quickly than non-PALS residents or those with ≤1 PGY (Multimedia Appendix 4). We observed no difference with previous resuscitation experience (data not shown).

Table 1. Participants’ demographics and clinical characteristics.
Demographics and clinical characteristicsRandomization Arm

ARa Glasses (n=10)PALSb pocket cards (n=10)
Age in years, mean (SD)27.9 (2.6)29.2 (2.6)
Sex (female), n (%)9 (90)6 (60)
Years of residency, mean (SD)2.4 (1.7)2.4 (1.7)
Number of residents having been enrolled in > 5 resuscitations in the past, n (%)4 (40)3 (30)
Number of PALS providers among residents, n (%)6 (60)7 (70)
Number of BLSc providers among residents, n (%)10 (100)10 (100)
Level of self-confidence in following AHA guidelines (on a scale of 1 to 5), mean (SD)1.8 (1.2)2.4 (0.8)

aAR: Augmented Reality.

bPALS: Pediatric Advanced Life Support.

cBLS: Basic Life Support.

Figure 4. Kaplan–Meier curves of proportion of pediatric residents using augmented reality glasses or conventional Pediatric Advanced Life Support pocket cards who a) initiated chest compression, or b) delivered first defibrillation shock during a simulated pulseless ventricular tachycardiapVT scenario (Log-rank test statistic, P=.81 and P=.99).
View this figure
Table 2. Mean time to resuscitation critical endpoints.
Outcomes

ARa Glasses
Mean (SD, 95% CI)
PALSb pocket cards
Mean (SD, 95% CI)
Time differencecP value
Time to initiation of CPRd28.0 (22.9, 11.6-44.4)25.6 (17.8, 12.9-38.3)2.4.80
Time to 1st defibrillation attempt146.2 (43.5, 115.1-177.3)145.7 (75.1, 92.0-199.4)0.5.99
Time to 2nd defibrillation attempt264.0 (73.9, 211.1-316.9)263.0 (74.2, 210.0-316.0)1.0.98
Time to epinephrine317.3 (62.6, 265.0-369.5)295.8 (97.7, 220.7-370.9)21.5.59
Time to 3rd defibrillation attempt396.6 (93.6, 329.7-463.5)389.0 (80.0, 314.8-462.9)7.7.86
Time to amiodarone450.1 (53.6, 408.9-491.3)492.7 (106.5, 416.5-568.9)42.6.28
Time to 4th defibrillation attempt542.8 (83.3, 478.7-606.8)526.8 (93.4, 455.0-598.6)16.0.71

aAR: Augmented Reality.

bPALS: Pediatric Advanced Life Support.

cTime difference represents absolute time difference between PALS pocket cards and AR Glasses.

dCPR: cardiopulmonary resuscitation.

Table 3. Errors and deviations from the American Heart Association’s pulseless ventricular tachycardia algorithm.
Critical Resuscitation Endpoints
(AHA pVTa algorithm)
AHAb recommended doses
(Joules per kg), (mL per kg)
Randomization Arm
ARc Glasses
Mean (SD, 95% CI)
% deviation
(P value)
PALSd ref. cards
Mean (SD, 95% CI)
% deviation
(P value)i
1st defibrillation attempt2.002.12 (0.71, 1.61-2.63)6 (.60)3.46 (2.23, 1.87-5.05)73 (.068)
2nd defibrillation attempt4.003.40 (0.97, 2.71-4.09)15 (.081)4.52 (1.53, 3.42-5.62)13 (.31)
Epinephrine 1:10,0000.10.1 (0, 0.1-0.1)0 (1.00)0.1 (0, 0.1-0.1)0 (1.00)
3rd defibrillation attempt4.004.00 (0, 4.00-4.00)0 (1.00)5.00 (1.51, 3.74-6.26)25 (.10)
Amiodarone0.10.1 (0, 0.1-0.1)0 (1.00)0.1 (0, 0.1-0.1)0 (1.00)
4th defibrillation attempt4.004.00 (0, 4.00-4.00)0 (1.00)5.68 (1.18, 4.84-6.52)42 (.0014)
Cumulative defibrillation dose14.013.52 (0.97, 12.32-14.72)3.4 (.10)18.66 (2.03, 13.87-23.45)e33.3 (.025)
Correct AHA sequence60%
40%f
Correct number of shocks80%
70%
Shock overdoses (>100 J)1/40 opportunities
21/40 opportunitiesg
Total number of errors5/40 opportunities
26/40 opportunitiesh
Defibrillation errors at 1st attempt, n (details)2 (40% lower, 200% higher)
5 (0.6 Joules per kg to 6 Joules per kg)
Defibrillation errors at 2nd attempt, n (details)3 (2 Joules per kg instead of 4 Joules per kg)
5 (1.2 Joules per kg to 6 Joules per kg)
Defibrillation errors at 3rd attempt, n (details)0
8 (2 Joules per kg to 6 Joules per kg)
Defibrillation errors at 4th attempt, n (details)0
8 (4.8 Joules per kg to 8 Joules per kg)

apVT: Pulseless ventricular tachycardia.

bAHA: American Heart Association.

cAR: Augmented Reality.

dPALS: Pediatric Advanced Life Support.

eDifference between AR glasses and PALS reference cards groups: P=.0010.

fDifference between AR glasses and PALS reference cards groups: P=.66.

gDifference between AR glasses and PALS reference cards groups: P<.001.

hDifference between AR glasses and PALS reference cards groups: P<.001.

i% deviation denotes percentage deviation from AHA recommended dose.

Errors and Deviations From AHA pVT Algorithm

Errors and deviations from the AHA pVT algorithm are summarized in Table 3. Out of 40 opportunities, 5 errors in defibrillation doses (12.5%) were committed during the whole scenario in allocation group A. This compares to 26 errors in defibrillation doses (65%) during the whole scenario in allocation group B (P<.001). Out of 10, one resident (10%) in allocation group B wrongly used a synchronized shock, 2/10 (20%) never delivered a third defibrillation, and 2/10 (20%) stopped the compressions after the 1st or 2nd defibrillation attempts.

The entire pVT algorithm was followed correctly in a stepwise fashion until ROSC by 60% of residents in allocation group A and 40% in allocation group B (P=.66).

Questionnaire About Perceived Stress and Satisfaction

The questionnaire was completed and returned by 100% of the participants. Participants in allocation group A and B rated the overall perceived stress to be 6.2 (95% CI 4.7-7.7) and 7.0 (95% CI 5.7-8.3), respectively on the Likert scale (P=.38). The usability, acceptance, and perception of the AR glasses were a major concern in our study and were assessed using a 17-item Unified Theory of Acceptance and Use of Technology (UTAUT) questionnaire [34]. The results will be published in a separate upcoming study.


Principal Findings

To our knowledge, this is the first randomized controlled trial to investigate the benefit of a wearable technology to improve pediatric residents’ performance and adherence with regard to AHA resuscitation guidelines. Using AR glasses, we found that time to first defibrillation attempt, time to other critical resuscitation endpoints, and drug dose delivery were not improved in terms of adherence to AHA guidelines. However, errors and deviations from the pVT guideline in terms of defibrillation doses and cumulative defibrillation doses were significantly reduced when compared with the use of the PALS pocket reference cards.

During resuscitation, time is a decisive success criterion. During the first 15 minutes, survival and favorable neurological outcome decrease linearly by 2.1% and 1.2% per minute, respectively [35]. Delays in initiating CPR have a detrimental effect on patient outcome regardless of the quality of resuscitation [36]. AHA therefore recommends pulseless patients of any age to receive immediate CPR without delay starting with chest compressions followed by a defibrillation within 180 seconds of a shockable rhythm. However, management, procedural skills, and adherence to these guidelines have been shown to fade after a few months of initial training [37-39]. With a critical patient’s condition and stress, physicians do not always have enough time to apply these guidelines and are prone to deviate from them [40]. PALS pocket cards are intended to resolve this problem by delivering fast and accurate summarized resuscitative knowledge and skills to providers. Nevertheless, Hunt et al have observed that despite availability of these recommendations, 66% of pediatric residents failed to start compressions within 60 seconds from the onset of a simulated pVT, 33% never started compressions, only 54% successfully defibrillated within 180 seconds, and 7% never discharged the defibrillator [40]. Similar results were obtained by Labrosse et al during a simulated pulseless shockable arrest scenario, where 25% of pediatric residents failed to start compressions and 4% never defibrillated a patient [41]. A more recent study among first-year pediatric residents showed a median time for initiation of CPR of 50 seconds and to first defibrillation of 282 seconds [42]. Pediatric residents performed better in our study with delays closer to AHA recommendations with a mean time to initiate chest compressions of 25 to 28 seconds and to first defibrillation of 146 seconds. In both allocation groups, 90% of residents started compressions within 60 seconds from the onset of pVT, and 70% defibrillated within 180 seconds. However, there were no advantages for residents to wear the AR glasses as they performed similarly with or without them regarding delay to critical resuscitation endpoints, whether they were PALS-certified or not. Despite an ergonomic-driven approach to adapt AHA resuscitation algorithms in AR glasses and a prior 15-minute training session for their use, our system failed to improve resuscitation efficiency in terms of time to major endpoints. An explanation might be that reducing further time to defibrillations and drug delivery was not achievable by residents training in emergency medicine. Indeed, our results in accordance with those from Hunt et al [40] showed a trend toward improvement over PGY of training and PALS certification in the mean time to all defibrillation attempts. It would be interesting in further studies to assess this assumption with certified emergency physicians.

Current AHA resuscitation guidelines emphasize 2 minutes of chest compressions between defibrillations attempts as optimal care for persistent pVT or VF in children [30,43]. In this study, residents performed similarly on average, with or without AR glasses and close to AHA recommendations.

Prompt defibrillation is crucial for termination of VF or pVT to achieve ROSC [43]. The AHA 2015 guidelines recommend treating pVT or VF in children with an initial dose of 2 J/kg [30]. For subsequent shocks, a dose of 4 J/kg is recommended, though higher energy levels may be considered up to adult dose, if not exceeding 10 J/kg (Figure 1). In this trial, residents using the PALS pocket cards were more prone to deviate from defibrillation doses than those using the AR glasses. On average, the shocks they provided were delivered with defibrillation doses 13% to 73% above AHA recommendations. “High” defibrillation doses concerned mostly the initial shock, with doses reaching up to 6 J/kg in 40% of cases when delivered by residents not wearing the glasses. In an observational study of 285 pediatric IHCA, a higher initial shock dose of more than 2 J/kg was not associated with superior termination of pVT or VF or improved survival rates [44]. In addition, children who were defibrillated with higher initial shock doses in the >3-5 J/kg range were significantly less likely to have termination of pVT or VF with ROSC or to survive the event. In our study, the final cumulative defibrillation dose delivered by the residents in the PALS pocket cards group was on average 33% significantly higher than the AHA expected value. This deviation was 10 times greater than that seen with residents wearing the AR glasses. In particular, 50% of residents using the PALS pocket cards used wrong energy doses whether it was for first or second shock delivery. In a total of 40 defibrillation attempts, they used wrong doses in 65% of cases for a cumulative defibrillation dose of 18.7 J/Kg. These errors were reduced by 53% and cumulative dose by 37% by using the AR glasses, suggesting a limited but worthwhile benefit of their use in simulated resuscitation. It would be interesting in further studies to determine whether this would translate into fewer errors in shock doses in real life.

Finally, in terms of drug dose concentrations, both groups in this study accurately administered epinephrine and amiodarone. The entire pVT algorithm was followed correctly until ROSC in a stepwise fashion by 60% of residents wearing AR glasses, compared with 40% of residents with the PALS pocket cards.

Limitations

This study has some limitations. First, it was conducted in a simulation-based resuscitation scenario. This choice was related to the ethical and organizational difficulties of conducting studies in real-life critical conditions. However, several studies have demonstrated the benefit of simulation as an investigative research methodology to answer research questions that otherwise could not be answered during CPR [45]. High-fidelity simulation is recognized as an essential tool to study resuscitation skills or technologies. Till date, none of the results obtained from simulation-based CPR studies disagreed with those obtained from studies in real life, confirming our study design choice. Realism was achieved, reflected by the stress levels experienced by the participants. They quoted the simulation as highly stressful when compared with real CPR situations.

Second, one might raise questions of our choice to choose Google Glass as the best AR glasses to display AHA guidelines. Indeed, despite its remarked 2012 commercial unveiling, Google Glass never reached its public audience and was discontinued in January 2015. However, there is a growing interest in recent literature toward its use in specialized medical fields [21,22,25,46,47]. We hypothesize that other AR glasses would not drastically change the results that we found with Google Glass, as their small size remains a major limitation to their use in displaying CPR algorithms. Further studies would be valuable to assess this assumption.

Conclusions

Taken together, our results support the interpretation that residents are not accurately following AHA recommendations during pediatric resuscitation, whether they are PALS certified or not. A wearable technology such as AR glasses might partially fill this gap and benefit patients by improving adherence and performance of residents to meet resuscitation requirements set by AHA, especially regarding delivery of defibrillation doses. In this sense, AR glasses appear as an interesting tool for emergency medicine and future studies are required to further examine this new paradigm.

Acknowledgments

We thank all the study participants for their contributions. All authors declare that there were no relationships or activities that could appear to have influenced the submitted work.

Authors' Contributions

JNS and SM did the literature search and reading, and JNS wrote the manuscript and prepared the figures and tables. JNS, FE, KH, PS, AS, and SM collected data, while JNS carried out the statistical analysis. JNS, FE, CL, AG, and SM carried out the development of the project software, and FE, LL, KH, CL, AG, SM, and JS were involved in the concept and design of the study and the critical review of manuscript content. All authors have contributed to, seen, and approved the final submitted version of the manuscript, had full access to all the data (including statistical reports and tables) in the study, and can take responsibility for the integrity of the data and the accuracy of the data analysis. The corresponding author confirms that he had full access to the participants’ data and endorsed the final responsibility for the submission. He further affirms that the manuscript is an honest, accurate, and transparent account of the study being reported, that no important aspects of the study have been omitted, and that any deviations from the study plan have been explained. Consent for publication was obtained from the participants.

Conflicts of Interest

None declared.

Multimedia Appendix 1

CONSORT-EHEALTH V1.6 checklist.

PDF File (Adobe PDF File), 740KB

Multimedia Appendix 2

Use of AR glasses to guide resuscitation.

JPG File, 5MB

Multimedia Appendix 3

Following the pVT pathway of the AHA algorithm when adapted for augmented reality glasses during a simulation-based pediatric cardiac arrest scenario.

JPG File, 6MB

Multimedia Appendix 4

The table details the mean time from onset of pVT to the study main endpoints in both PALS and non-PALS residents.

PNG File, 512KB

  1. Thomas L, Cullum N, McColl E, Rousseau N, Soutter J, Steen N. Guidelines in professions allied to medicine. Cochrane Database Syst Rev 2000(2):CD000349. [CrossRef] [Medline]
  2. -. Standards and guidelines for Cardiopulmonary Resuscitation (CPR) and Emergency Cardiac Care (ECC). National Academy of Sciences - National Research Council. JAMA 1986 Jun 06;255(21):2905-2989. [Medline]
  3. Hazinski MF, Nolan JP, Aickin R, Bhanji F, Billi JE, Callaway CW, et al. Part 1: Executive Summary: 2015 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations. Circulation 2015 Oct 20;132(16 Suppl 1):S2-39 [FREE Full text] [CrossRef] [Medline]
  4. Meaney PA, Bobrow BJ, Mancini ME, Christenson J, de Caen AR, Bhanji F, CPR Quality Summit Investigators, the American Heart Association Emergency Cardiovascular Care Committee, the Council on Cardiopulmonary, Critical Care, Perioperative and Resuscitation.. Cardiopulmonary resuscitation quality: [corrected] improving cardiac resuscitation outcomes both inside and outside the hospital: a consensus statement from the American Heart Association. Circulation 2013 Jul 23;128(4):417-435 [FREE Full text] [CrossRef] [Medline]
  5. CDC. CDC. 2011. National hospital ambulatory medical care survey: 2011 emergency department summary tables   URL: https://www.cdc.gov/nchs/data/ahcd/nhamcs_emergency/2011_ed_web_tables.pdf [accessed 2017-01-22] [WebCite Cache]
  6. Knudson JD, Neish SR, Cabrera AG, Lowry AW, Shamszad P, Morales DL, et al. Prevalence and outcomes of pediatric in-hospital cardiopulmonary resuscitation in the United States: an analysis of the Kids' Inpatient Database*. Crit Care Med 2012 Nov;40(11):2940-2944. [CrossRef] [Medline]
  7. Topjian AA, Berg RA, Nadkarni VM. Pediatric cardiopulmonary resuscitation: advances in science, techniques, and outcomes. Pediatrics 2008 Nov;122(5):1086-1098 [FREE Full text] [CrossRef] [Medline]
  8. Gerein RB, Osmond MH, Stiell IG, Nesbitt LP, Burns S. What are the etiology and epidemiology of out-of-hospital pediatric cardiopulmonary arrest in Ontario, Canada? Acad Emerg Med 2006 Jun;13(6):653-658 [FREE Full text] [CrossRef] [Medline]
  9. Kuisma M, Suominen P, Korpela R. Paediatric out-of-hospital cardiac arrests--epidemiology and outcome. Resuscitation 1995 Oct;30(2):141-150. [Medline]
  10. Sirbaugh PE, Pepe PE, Shook JE, Kimball KT, Goldman MJ, Ward MA, et al. A prospective, population-based study of the demographics, epidemiology, management, and outcome of out-of-hospital pediatric cardiopulmonary arrest. Ann Emerg Med 1999 Feb;33(2):174-184. [Medline]
  11. Atkins DL, Everson-Stewart S, Sears GK, Daya M, Osmond MH, Warden CR, Resuscitation Outcomes Consortium Investigators. Epidemiology and outcomes from out-of-hospital cardiac arrest in children: the resuscitation outcomes consortium epistry-cardiac arrest. Circulation 2009 Mar 24;119(11):1484-1491 [FREE Full text] [CrossRef] [Medline]
  12. McEvoy MD, Field LC, Moore HE, Smalley JC, Nietert PJ, Scarbrough SH. The effect of adherence to ACLS protocols on survival of event in the setting of in-hospital cardiac arrest. Resuscitation 2014 Jan;85(1):82-87 [FREE Full text] [CrossRef] [Medline]
  13. Girotra S, Nallamothu BK, Spertus JA, Li Y, Krumholz HM, Chan PS, American Heart Association Get with the Guidelines–Resuscitation Investigators. Trends in survival after in-hospital cardiac arrest. N Engl J Med 2012 Nov 15;367(20):1912-1920 [FREE Full text] [CrossRef] [Medline]
  14. Hunt EA, Duval-Arnould JM, Nelson-McMillan KL, Bradshaw JH, Diener-West M, Perretta JS, et al. Pediatric resident resuscitation skills improve after “rapid cycle deliberate practice” training. Resuscitation 2014 Jul;85(7):945-951. [CrossRef] [Medline]
  15. Kämäräinen A. Out-of-hospital cardiac arrests in children. J Emerg Trauma Shock 2010 Jul;3(3):273-276 [FREE Full text] [CrossRef] [Medline]
  16. Jayaram N, McNally B, Tang F, Chan PS. Survival after out-of-hospital cardiac arrest in children. J Am Heart Assoc 2015 Oct 08;4(10):e002122 [FREE Full text] [CrossRef] [Medline]
  17. Field LC, McEvoy MD, Smalley JC, Clark CA, McEvoy MB, Rieke H, et al. Use of an electronic decision support tool improves management of simulated in-hospital cardiac arrest. Resuscitation 2014 Jan;85(1):138-142 [FREE Full text] [CrossRef] [Medline]
  18. Low D, Clark N, Soar J, Padkin A, Stoneham A, Perkins GD, et al. A randomised control trial to determine if use of the iResus© application on a smart phone improves the performance of an advanced life support provider in a simulated medical emergency. Anaesthesia 2011 Apr;66(4):255-262 [FREE Full text] [CrossRef] [Medline]
  19. Merchant RM, Abella BS, Abotsi EJ, Smith TM, Long JA, Trudeau ME, et al. Cell phone cardiopulmonary resuscitation: audio instructions when needed by lay rescuers: a randomized, controlled trial. Ann Emerg Med 2010 Jun;55(6):538-543.e1. [CrossRef] [Medline]
  20. Semeraro F, Frisoli A, Loconsole C, Bannò F, Tammaro G, Imbriaco G, et al. Motion detection technology as a tool for cardiopulmonary resuscitation (CPR) quality training: a randomised crossover mannequin pilot study. Resuscitation 2013 Apr;84(4):501-507. [CrossRef] [Medline]
  21. Chaballout B, Molloy M, Vaughn J, Brisson IR, Shaw R. Feasibility of augmented reality in clinical simulations: using google glass with manikins. JMIR Med Educ 2016 Mar 07;2(1):e2 [FREE Full text] [CrossRef] [Medline]
  22. Chai PR, Wu RY, Ranney ML, Bird J, Chai S, Zink B, et al. Feasibility and acceptability of google glass for emergency department dermatology consultations. JAMA Dermatol 2015 Jul;151(7):794-796. [CrossRef] [Medline]
  23. Iqbal MH, Aydin A, Lowdon A, Ahmed HI, Muir GH, Khan MS, et al. The effectiveness of google glass as a vital signs monitor in surgery: ac simulation study. Int J Surg 2016 Dec;36(Pt A):293-297. [CrossRef] [Medline]
  24. Liebert CA, Zayed MA, Aalami O, Tran J, Lau JN. Novel use of google glass for procedural wireless vital sign monitoring. Surg Innov 2016 Aug;23(4):366-373. [CrossRef] [Medline]
  25. Mitrasinovic S, Camacho E, Trivedi N, Logan J, Campbell C, Zilinyi R, et al. Clinical and surgical applications of smart glasses. Technol Health Care 2015;23(4):381-401. [CrossRef] [Medline]
  26. Stetler J, Resendes E, Martinez-Parachini JR, Patel K, Amsavelu S, Tarar MN, et al. Hands-free zoom and pan technology improves the accuracy of remote electrocardiogram interpretation using google glass. Int J Cardiol 2016 Feb 01;204:147-148. [CrossRef] [Medline]
  27. Ehrler F, Siebert J, Haddad K, Sahin A, Schrurs P, Diener R, et al. Adapting guidelines for google glass: the case of pediatric CPR. Stud Health Technol Inform 2016;224:141-145. [Medline]
  28. Nadkarni VM, Larkin GL, Peberdy MA, Carey SM, Kaye W, Mancini ME, National Registry of Cardiopulmonary Resuscitation Investigators. First documented rhythm and clinical outcome from in-hospital cardiac arrest among children and adults. JAMA 2006 Jan 4;295(1):50-57. [CrossRef] [Medline]
  29. Samson RA, Nadkarni VM, Meaney PA, Carey SM, Berg MD, Berg RA, American Heart Association National Registry of CPR Investigators. Outcomes of in-hospital ventricular fibrillation in children. N Engl J Med 2006 Jun 01;354(22):2328-2339. [CrossRef] [Medline]
  30. de Caen AR, Berg MD, Chameides L, Gooden CK, Hickey RW, Scott HF, et al. Part 12: pediatric advanced life support: 2015 American Heart Association guidelines update for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 2015 Nov 3;132(18 Suppl 2):S526-S542 [FREE Full text] [CrossRef] [Medline]
  31. Chan PS, Krumholz HM, Nichol G, Nallamothu BK, American Heart Association National Registry of Cardiopulmonary Resuscitation Investigators. Delayed time to defibrillation after in-hospital cardiac arrest. N Engl J Med 2008 Jan 03;358(1):9-17. [CrossRef] [Medline]
  32. Dilley SJ, Weiland TJ, O'Brien R, Cunningham NJ, Van Dijk JE, Mahoney RM, et al. Use of a checklist during observation of a simulated cardiac arrest scenario does not improve time to CPR and defibrillation over observation alone for subsequent scenarios. Teach Learn Med 2015;27(1):71-79. [CrossRef] [Medline]
  33. Sealed envelope.   URL: https://www.sealedenvelope.com [accessed 2017-01-21] [WebCite Cache]
  34. JSTOR. MIS Quarterly   URL: http://www.jstor.org/stable/30036540 [accessed 2017-03-29] [WebCite Cache]
  35. Matos RI, Watson RS, Nadkarni VM, Huang H, Berg RA, Meaney PA, American Heart Association’s Get With The Guidelines–Resuscitation (Formerly the National Registry of Cardiopulmonary Resuscitation) Investigators. Duration of cardiopulmonary resuscitation and illness category impact survival and neurologic outcomes for in-hospital pediatric cardiac arrests. Circulation 2013 Jan 29;127(4):442-451. [CrossRef] [Medline]
  36. Ross JC, Trainor JL, Eppich WJ, Adler MD. Impact of simulation training on time to initiation of cardiopulmonary resuscitation for first-year pediatrics residents. J Grad Med Educ 2013 Dec;5(4):613-619 [FREE Full text] [CrossRef] [Medline]
  37. Smith KK, Gilcreast D, Pierce K. Evaluation of staff's retention of ACLS and BLS skills. Resuscitation 2008 Jul;78(1):59-65. [CrossRef] [Medline]
  38. Tarantinos K, Chalkias A, Giotakis E, Athanasopoulou P, Xanthos T. Retention of knowledge and skills after advanced cardiovascular life support courses. Am J Emerg Med 2014 Sep;32(9):1143-1147. [CrossRef] [Medline]
  39. Yang C, Yen Z, McGowan JE, Chen HC, Chiang W, Mancini ME, et al. A systematic review of retention of adult advanced life support knowledge and skills in healthcare providers. Resuscitation 2012 Sep;83(9):1055-1060. [CrossRef] [Medline]
  40. Hunt EA, Vera K, Diener-West M, Haggerty JA, Nelson KL, Shaffner DH, et al. Delays and errors in cardiopulmonary resuscitation and defibrillation by pediatric residents during simulated cardiopulmonary arrests. Resuscitation 2009 Jul;80(7):819-825. [CrossRef] [Medline]
  41. Labrosse M, Levy A, Donoghue A, Gravel J. Delays and errors among pediatric residents during simulated resuscitation scenarios using pediatric advanced life support (PALS) algorithms. Am J Emerg Med 2015 Oct;33(10):1516-1518. [CrossRef] [Medline]
  42. Jeffers J, Eppich W, Trainor J, Mobley B, Adler M. Development and evaluation of a learning intervention targeting first-year resident defibrillation skills. Pediatr Emerg Care 2016 Apr;32(4):210-216. [CrossRef] [Medline]
  43. Kleinman ME, Chameides L, Schexnayder SM, Samson RA, Hazinski MF, Atkins DL, et al. Part 14: pediatric advanced life support: 2010 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 2010 Nov 02;122(18 Suppl 3):S876-S908 [FREE Full text] [CrossRef] [Medline]
  44. Meaney PA, Nadkarni VM, Atkins DL, Berg MD, Samson RA, Hazinski MF, American Heart Association National Registry of Cardiopulmonary Resuscitation Investigators. Effect of defibrillation energy dose during in-hospital pediatric cardiac arrest. Pediatrics 2011 Jan;127(1):e16-e23. [CrossRef] [Medline]
  45. Cheng A, Auerbach M, Hunt EA, Chang TP, Pusic M, Nadkarni V, et al. Designing and conducting simulation-based research. Pediatrics 2014 Jun;133(6):1091-1101 [FREE Full text] [CrossRef] [Medline]
  46. Chang JY, Tsui LY, Yeung KS, Yip SW, Leung GK. Surgical vision: google glass and surgery. Surg Innov 2016 Aug;23(4):422-426. [CrossRef] [Medline]
  47. Aungst TD, Lewis TL. Potential uses of wearable technology in medicine: lessons learnt from Google Glass. Int J Clin Pract 2015 Oct;69(10):1179-1183. [CrossRef] [Medline]


AHA: American Heart Association
AR: Augmented Reality
CA: Cardiac Arrest
CPR: Cardiopulmonary Resuscitation
IHCA: In-Hospital Cardiac Arrest
OHCA: Out-of-Hospital Cardiac Arrest
PALS: Pediatric Advanced Life Support
PED: Pediatric Emergency Department
PGY: Post-Graduation Years
pVT: pulseless ventricular tachycardia
ROSC: Return Of Spontaneous Circulation
VF: Ventricular Fibrillation


Edited by G Eysenbach; submitted 21.01.17; peer-reviewed by K Morawski, M Leary, A Elmessiry; comments to author 26.03.17; revised version received 03.04.17; accepted 28.04.17; published 29.05.17

Copyright

©Johan N Siebert, Frederic Ehrler, Alain Gervaix, Kevin Haddad, Laurence Lacroix, Philippe Schrurs, Ayhan Sahin, Christian Lovis, Sergio Manzano. Originally published in the Journal of Medical Internet Research (http://www.jmir.org), 29.05.2017.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research, is properly cited. The complete bibliographic information, a link to the original publication on http://www.jmir.org/, as well as this copyright and license information must be included.