TY - JOUR AU - Mazoteras-Pardo, Victoria AU - Becerro-De-Bengoa-Vallejo, Ricardo AU - Losa-Iglesias, Marta Elena AU - Martínez-Jiménez, Eva María AU - Calvo-Lobo, César AU - Romero-Morales, Carlos AU - López-López, Daniel AU - Palomo-López, Patricia PY - 2020 DA - 2020/7/24 TI - QardioArm Blood Pressure Monitoring in a Population With Type 2 Diabetes: Validation Study JO - J Med Internet Res SP - e19781 VL - 22 IS - 7 KW - blood pressure KW - hypertension KW - type 2 diabetes KW - mobile applications KW - software validation AB - Background: Home blood pressure monitoring has many benefits, even more so, in populations prone to high blood pressure, such as persons with diabetes. Objective: The purpose of this research was to validate the QardioArm mobile device in a sample of individuals with noninsulin-dependent type 2 diabetes in accordance with the guidelines of the second International Protocol of the European Society of Hypertension. Methods: The sample consisted of 33 patients with type 2 diabetes. To evaluate the validity of QardioArm by comparing its data with that obtained with a digital sphygmomanometer (Omron M3 Intellisense), two nurses collected diastolic blood pressure, systolic blood pressure, and heart rate with both devices. Results: The analysis indicated that the test device QardioArm met all the validation requirements using a sample population with type 2 diabetes. Conclusions: This paper reports the first validation of QardioArm in a population of individuals with noninsulin-dependent type 2 diabetes. QardioArm for home monitoring of blood pressure and heart rate met the requirements of the second International Protocol of the European Society of Hypertension. SN - 1438-8871 UR - http://www.jmir.org/2020/7/e19781/ UR - https://doi.org/10.2196/19781 UR - http://www.ncbi.nlm.nih.gov/pubmed/32706672 DO - 10.2196/19781 ID - info:doi/10.2196/19781 ER -